General
Load Capability
Maximum Isolation Voltage
Temperature Range
Relative Humidity
Altitude
Vibration
Shock
Power Requirements
Weight
Dimensions
Protection Safety Class 1
1-4
Table 1-2. Specifications (cont.)
Drives all resistive and capacitive loads consistent with current and
compliance voltage capability. For inductive loads greater than
1 H and less than maximum inductive loads, output current is
limited to 16 A rms (22.6 A peak)
20 V dc or 20 V ac rms
Operating: 0 C to 50 C
Storage: +20 C to 56 C
50 % to 50 C, 75 % to 40 C, 95 % to 25 C
Operating: 0 ft to10,000 ft
Non-operating: 0 ft to 40,000 ft
2 g maximum, 5 Hz to 55 Hz for 15 minutes
15 g maximum, half sinewave
100, 110, 115, 120, 200, 220, 230, or 240 V ac 10 %, switch-
selectable, 50 Hz to 60 Hz, 300 W
27 kg (50 lb)
17.8 cm H x 43.2 cm W x 55.9 cm D (7 in H x 17 in W x 22 in D),
case only (See Figure 1-1).
Relates solely to insulation or grounding properties as defined in
IEC 348.