Fluke 5220A Instruction Manual page 15

Transconductance amplifier
Table of Contents

Advertisement

Electromagnetic Compatibility
DC Mode (including 5100 Series B)
Output Range
Accuracy of Output
Resolution
Temperature Coefficient
Line Regulation
Load Regulation
AC Mode (including 5100 Series B)
Output Range
Accuracy of Output
Resolution
Temperature Coefficient
Short Term Stability
Harmonic Distortion and Noise
Line Regulation
Load Regulation
Table 1-2. Specifications (cont.)
This instrument is designed to operate in Standard Laboratory
environments where in electromagnetic environment is highly
controlled. If used in areas with RF fields 0.4 V/m there could be
errors in measurement.
1 to 19.9999 A
(0.025 % of selected output + 1 mA)
0.1 mA
(0.003 % of selected output + 100 A) / C, above 30 C and
below 20 C
Output changes less than 0.001 % for a 10 % change in line
voltage.
Output changes less than (0.005 % + 0.3 mA) for a full load
changes of 4 V of compliance.
1 A rms to 19.9999 A rms
(0.07 % of selected output + 1 mA rms) from 50 Hz to 1 kHz,
and (0.07 % of selected output + 1 mA rms) x f from 1 kHz to 5
kHz, where f = frequency in kHz.
0.1 mA rms
(0.003 % of selected output + 100 A rms) per C, above 30 C
and below 20 C.
Output changes less than (0.02 % + 500 A rms) in 10 minutes,
with constant line, load, and temperature.
(0.07 % of output + 1 mA rms) over frequency range of 30 Hz to
1 kHz a noise bandwidth of 300 kHz, (0.07 % of output + 1 mA)
x f from 1 kHz to 5 kHz, where f = frequency in Hz.
Output changes less than 0.005 % for 10 % changes in line
voltage.
For frequencies less than 1 kHz:
(0.005 % lout + 150 A) per volt of output compliance voltage.
For frequencies greater than 1 kHz:
(0.005 % lout + 150 A) x f per volt of output compliance
voltage, where f = frequency in kHz.
Introduction and Specifications
Specifications
1
1-3