Loop Power - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

-200 to 100
0.07 C
100
Pt(3916)
100 to 630
0.02 % + 0.05 C
-200 to 100
0.08 C
100
Pt(3926)
100 to 630
0.02 % + 0.06 C
10
-100 to 260
0.2 C
Cu(427)
120
-80 to 260
0.1 C
Ni(672)
[1] Specifications are valid to k=3
Sensor inaccuracies not included
[2] For two and three-wire RTD measurements, add 0.4 C to the specifications.
Resolution: 0.01 C except 0.1 C for 10 Ω Cu(427)
Temperature Coefficient: 0.01 C/ C for measure, 0.02 C/ C (<18 C or >28 C) for source
[3] Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
RTD Reference:
Pt(385): IEC 60751, 2008
Pt(3916): JIS C 1604, 1981
Pt(3926), Cu(427), Ni(672): Minco Application Aid #18

Loop Power

Open Circuit
26 V 10 %
Short circuit protected to 25 mA
Output Resistance: 250
nominal
Temperature, RTDs
Degrees or % of Reading
0.14 C
1 mA
0.04 % + 0.10 C
0.16 C
1 mA
0.04 % + 0.12 C
0.4 C
3 mA
0.2 C
1 mA
18 V minimum at 22 mA
Documenting Process Calibrator
Detailed Specifications
[1]
0.05 C
0.10 C
0.0125 % + 0.04 C
0.025 % + 0.08 C
0.05 C
0.10 C
0.0125 % + 0.04 C
0.025 % + 0.08 C
0.2 C
0.4 C
0.04 C
0.08 C
Loaded Circuit
0.1 to 10 mA
0.1 to 10 mA
1 to 10 mA
0.1 to 10 mA
115

Advertisement

Table of Contents
loading

This manual is also suitable for:

754

Table of Contents