Epson USN 52R Operating Manual page 90

Lightweight, compact ultrasonic flaw detectors
Table of Contents

Advertisement

Configuring the USN 52R/USN 52L
VELOCITY #1
Preselecting fixed values for
VELOCITY #2
sound velocity
When calibrating the display range, you can select be-
tween two fixed sound velocity values in the coarse
mode.
In this function you determine which values are avail-
able for selection.
• You can set the sound velocity between 1000 m/s
and 9999 m/s (continuously adjustable).
Operation:
– Highlight the center function group in the left function
table with I.
– With M set the required values in VELOCITY #1
and VELOCITY #2.
H Note:
The preset default values here are as follows:
• 5918 m/s for longitudinal sound velocity
• 3353 m/s for transverse sound velocity
Krautkramer USN 52R/USN 52L
dB STEP Sets step range
You can set the gain with five different step ranges. In
this function you enter your own value for the adjust-
able step.
• You are able to enter a step range between 6.5 dB
und 24.0 dB in steps of 0.5 dB.
Operation:
– Highlight the center function group in the left function
table with I.
– With M set the required step range.
Issue 05, 02/00
Operation
5-45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usn 52l

Table of Contents