Automatic calibration

The functions for automatic calibration are in the "right"
function table:
AUTO CAL Switches on automatic calibration
GATE LOGIC Sets the gate for calibration
Description of automatic calibration can be found in
Chapter 5.
How to change values of the gate settings please refer
to chapter 5.2.
5-62
5.13 Setting the measurement
methods
MEASURE Method of measurement for
With this function you can select a method for the
sound path measurement in the gate range. You have
three possibilities:
• OFF: No measurement
• 0 TO 1st: Measurement from zero point to the first or
the highest echo above the threshold
• MULTECHO: Measurement between the first two
echoes or between the two highest echoes above the
threshold
Operation:
– Highlight the right function group in the left function
table using I.
– With M select the required setting in MEASURE.
Issue 05, 02/00

Setting the measurement methods

sound path measurement
Krautkramer USN 52R/USN 52L