Adjustment Of The Usn 52R/Usn 52L; Gain Gain Adjustment - Epson USN 52R Operating Manual

Lightweight, compact ultrasonic flaw detectors
Table of Contents

Advertisement

Adjustment of the USN 52 R
5.2 Adjustment of the USN 52R/
USN 52L
GAIN
Gain adjustment
With this function the echoes of the reflectors to be
detected are amplified so that they appear on the dis-
play at the required height.
In order that you are able to change the gain at any
time, you will find this function in all function groups of
the first operating level.
Operation:
– Change the setting of GAIN with M.
Krautkramer USN 52R/USN 52L
D
-Setting the dB STEP
You set the gain with a certain step size which you can
select with D. There are five steps available:
• 6.5 to 24 dB
(variable via dB STEP in the second operating level,
refer to Chapter 5.6)
• 6.0 dB
• 2.0 dB
• 1.0 dB
• 0.5 dB
• 0.1 dB
Operation:
• Press D in order to change between the five steps.
The set step is displayed as a value in the GAIN
function.
Issue 05, 02/00
Operation
5-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usn 52l

Table of Contents