Epson USN 52R Operating Manual page 145

Lightweight, compact ultrasonic flaw detectors
Table of Contents

Advertisement

Interfaces, Peripherals
between 1 and 6 for ##. $$ stands for the length of the
field, text for the field label. The length of the name
must not exceed the field length. Shorter names are
automatically left-justified. You have to stick to the se-
quence of the fields when defining field lengths and
names, that means start with HF 1 ..., continue with
HF 2 ..., etc.
The sum of all field lenghts must not exceed 80.
You can define active fields which are automatically
filled in with the current data from the A-scan. In that
case, these fields can no longer be selected for the
entry of texts when storing. To define active fields, use
the remote control codes of the corresponding functions
preceded by the character \. Type the name of the field
directly after the two-digit remote control code.
Example: <ESC>HF 4 11\PK AMPLITUDE<RETURN>
The field number 4 will be 11 characters long, it will
have the name AMPLITUDE and be automatically filled
in with the amplitude value.
The possible codes are:
PK (amplitude, length 9 characters)
TK (sound path, length 9 characters)
PR (projection distance, length 9 characters)
8-18
DP (depth, length 8 characters)
The field length is always determined, other values
(e.g. 11) are ignored.
Example for a header definition:
<ESC>HN 9<RETURN>
<ESC>HF 1 10 OPERATOR<RETURN>
<ESC>HF 2 7 DATA<RETURN>
<ESC>HF 3 10 X-COORD<RETURN>
<ESC>HF 4 10 Y-COORD<RETURN>
<ESC>HF 5 9\PK AMPL.<RETURN>
<ESC>HF 6 9\TK S-PATH<RETURN>
<ESC>HF 7 9\PR P-DIST<RETURN>
<ESC>HF 8 8\DP DEPTH<RETURN>
<ESC>HF 9 8 COMMENT<RETURN>
After the last field has been defined, the reply is:
OK.
The fields 5 to 8 are automatically filled in.
Issue 05, 02/00
Memo function
Krautkramer USN 52R/USN 52L

Advertisement

Table of Contents
loading

This manual is also suitable for:

Usn 52l

Table of Contents