Automatic Zero Maintenance; Warm-Up Time; Thermal Equilibrium; Power-Up - Datalogic MAGELLAN 8500 Product Reference Manual

Hide thumbs Also See for MAGELLAN 8500:
Table of Contents

Advertisement

Warm-Up Time

The two warm-up periods can be performed concurrently, thereby reducing
the total required warm-up time to 60 minutes.
NOTE
Product Reference Guide

Automatic Zero Maintenance

The scale's software constantly monitors and adjusts the Zero point as
long as the deviation is within acceptable limits
any debris accumulation or removal. During power-up, the scale automat-
ically re-zeros after verifying that all subsystems are functional. Addition-
ally, the scale may be manually "zeroed" by pushing the Scale Zero Push
Button located on the top of the vertical enclosure.
There are two pertinent warm-up times that apply to the scanner or scan-
ning-scale:

Thermal Equilibrium

When the unit is moved from a cooler temperature (such as a storage area)
to a warmer environment (such as a checkstand location), 60 minutes
must be allowed to acclimate the unit to ambient conditions prior to cali-
bration or operation.

Power-up

Once installed and powered up, a warm-up time of 15 minutes must be
allowed before calibrating or performing weighing operations.

User Configurable Warm-up

The user may configure the unit for a pre-programmed warm-up time
that is activated every time the scanner is powered up. During this time,
the scale is viewed by the POS terminal as off-line.
1. Acceptable limitis of deviation are set at -0.2 to +0.6 pounds (-0.078kg to 0.23kg), which is -0.67 to
+2.0% of total capacity.
1
, while compensating for
1-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Magellan 8500 omega

Table of Contents