Brady Bradyprinter 2461 Operator's Manual page 68

Hide thumbs Also See for Bradyprinter 2461:
Table of Contents

Advertisement

Step
Operator Action
Press the
6
Use the
scroll to 'Sensor
Calibration'.
Press the
Use the
7
scroll to 'Advanced Entry'.
Press the
Press the
8
9
Place the label under the
Sensor Eye Mark and lower
the printhead.
Using the
increment the Gain
Number. Press the
key to select the new setting
and then record the sensor
reading (the 'yyy' value).
Repeat this process for each
of the 16 Gain Numbers.
These are the Label Values.
(Continued next page)
58
Displayed Message
key.
key to
key.
key to
key.
key.
key
-OR-
Comment
See Section
4.1.1 for
details.
Press the
key to
abort this
procedure.
See Section
4.1.1 for
details.
See Section
4.1.1.
Ensure that an
unmarked label
area (without
pre-printing) is
in the Media
Sensor.