Performance Tests; Required Equipment - Fluke 772 Calibration Manual

Milliamp process clamp meter
Hide thumbs Also See for 772:
Table of Contents

Advertisement

Performance Tests

To prevent electrical shock, personal injury, or fire:
Repairs or Meter servicing must be done only by qualified
personnel.
Do not do the verification tests or calibration adjustment in
this manual unless qualified.
The tests that follow verify the functions of the Meter. If the Meter fails the verification
tests, repair is necessary. For Meter servicing, see "How to Contact Fluke".

Required Equipment

Required equipment for the performance tests is in Table 2. If the recommended models
are not available, equipment with equivalent specifications can be used.
Equipment
Calibrator
DMM
Lab Supply
Figure 2. Battery Replacement
XWWarning
Table 2. Required Equipment
Minimum Required Characteristics
DC milliamps:
0-24.00 mA = 0.073 %
24.0-100.0 mA = 0.375 %
DC Volts:
0-30.00 V = 0.267 %
DC Current:
0-24.00 mA = 0.375 %
DC Volts:
0-10 V = 0.1 %
6 Vdc -0.5 V
Milliamp Process Clamp Meter
Performance Tests
gep02.eps
Recommended Model
Fluke 55xxA
Fluke 88xxA
-
7