Additional Installation Information

Advanced OSD

Several control functions are available by entering the advanced OSD. These control functions are additional OSD menu items,
which are always accessible after opening the standard OSD.
Control Function
Description
EDID
Change the picture format to be recognized by the operating system
Input Source
Change the input source
Confi guration
Change luminance and gamma correction factory presets. The factory presets are:
The confi guration 1 parameter may be modifi ed anytime by a new display calibration.
NOTE: Regarding display calibration, please refer to the GammaCompMD QA Manual.
DICOM
Make DICOM conformance of the display by measuring and checking the display properties.
conformance test
How to change the EDID setting (Portrait / Landscape)
1. Click the (+) plus button or (-) minus button on the back side of the monitor once to display the "INFORMATION" OSD.
A display, which is similar to this picture, opens on the screen.
2. Hold down the (+) plus button and the (-) minus button simultaneously for over 2 seconds.
The "SWITCHING EDID" menu opens on the screen. Select "Portrait" or "Landscape" by the (+) plus button or the (-) minus
button.
Portrait format - 2048 x 2560 pixels [Factory Default]
Landscape format - 2560 x 2048 pixels
DVI
DisplayPort
Confi guration
Maximum
1
500
2
550
3
400
Luminance (cd/m
2
)
Minimum
Ambient
1.0
0.0
1.0
0.0
1.0
0.0
English-12
Gamma
Remark
DICOM
Default
DICOM
DICOM