How to Change Confi guration (Preset 1, 2, 3)
The NEC model MD211G5 is shipped with 3 factory preset confi gurations, as listed below.
Confi guration
Maximum
1
500
2
550
3
400
The confi guration can be changed via an OSD menu. The confi guration 1 parameters may be modifi ed anytime by a new
display calibration. Regarding display calibration, please refer to the GammaCompMD QA Manual.
1. Click the (+) plus button or (-) minus button on the back side of the monitor twice to display the "PRESENT STATE" OSD.
A display like this picture opens on the screen.
2. With the present operating state displayed, press and hold both the buttons for 2 seconds. The switching confi guration
screen appears.
With the screens above displayed, press a power-switch
and SWITCHING CONFIGURATION screen in that order.
3. Select your desired confi guration by the (+) plus button or the (-) minus button.
4. After selecting your preferred CONFIGURATION mode, hold down the (+) plus button and the (-) minus button
simultaneously for over 2 seconds again. Then confi rm any change of confi guration in the "PRESENT STATE" OSD.
5. Click the (+) plus button or (-) minus button once to close the "PRESENT STATE" OSD.
2
Luminance (cd/m
)
Minimum
Ambient
1.0
0.0
1.0
0.0
1.0
0.0
English-14
Gamma
DICOM
[Factory Default]
DICOM
DICOM
quickly, the OSD switches to SWITCHING INPUT SOURCE
Remark