Calibration Adjustments - Agilent Technologies U1241B User's Manual And Service Manual

Handheld digital multimeters
Hide thumbs Also See for U1241B:
Table of Contents

Advertisement

Calibration Adjustments

Review the
N O T E
adjustment procedures.
1 Turn the rotary switch to Test Function position, as shown in
2 After unsecuring the multimeter, the multimeter will go into the
3 The primary display will show the reference value of the Cal Items.
4 Use the
5 Apply the input signal as shown in the Valid Input Reference Value
Users are highly recommended to complete the adjustments in the same order as shown in
N O T E
Table
6 Enter the actual applied input, see
7 Press
8 Upon completion of each adjustment value, the secondary display will
If the adjustment failed, check the input value, range, function, and entered adjustment
N O T E
value and repeat the adjustment steps.
U1241B/U1242B User's and Service Guide
Test Considerations
Table 5-
3.
adjustment mode, see
Configure each Cal Item shown in Valid Adjustment Input Reference
Values in
Table 5-
3.
and
column of the
Table 5-
There is no bar graph display for temperature adjustment.
5-3.
to start the adjustments. The CAL flashes in the secondary
display indicating the calibration is in progress.
show PAS. If the adjustment failed, the multimeter will sound a long
beep and the calibration error number appears in the secondary
display. The primary display remains at the current Cal Item.
Performance Tests and Calibration
and
Adjustment Considerations
Unsecuring the Multimeter for
arrow keys to select the Cal Range.
3. The bar graph will display the input reading.
Entering Adjustment
before beginning the
Calibration.
Values.
5
51