Table of Contents

Advertisement

Operating Manual
BD-S / BD-S-UL (E1)
ED-S / ED-S-UL (E1)
FD-S / FD-S-UL (E1)
with R-S microprocessor temperature controller
Model
Model version
BD-S 56
BDS056-230V
BD-S-UL 56
BDS056UL-120V
BD-S 115
BDS115-230V
BD-S-UL 115
BDS115UL-120V
ED-S 56
EDS056-230V
ED-S-UL 56
EDS056UL-120V
ED-S 115
EDS115-230V
ED-S-UL 115
EDS115UL-120V
FD-S 56
FDS056-230V
FD-S-UL 56
FDS056UL-120V
FD-S 115
FDS115-230V
FD-S-UL 115
FDS115UL-120V
BINDER GmbH
 Address: Post office box 102, 78502 Tuttlingen, Germany  Phone: +49 7462 2005 0
 Fax: +49 7462 2005 100  Internet: http://www.binder-world.com
 E-mail: info@binder-world.com  Service Hotline: +49 7462 2005 555
 Service Fax: +49 7462 2005 93 555  Service E-Mail: service@binder-world.com
 Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
 Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
 Service Hotline Russia and CIS: +7 495 988 15 16
Issue 05/2019
Incubators with natural convection
Drying and heating ovens with natural
convection
Drying and heating ovens with forced
convection
Art. No.
9090-0016, 9190-0016
9090-0017, 9190-0017
9090-0022, 9190-0022
9090-0023, 9190-0023
9090-0014, 9190-0014
9090-0015, 9190-0015
9090-0020, 9190-0020
9090-0021, 9190-0021
9090-0018, 9190-0018
9090-0019, 9190-0019
9090-0024, 9190-0024
9090-0025, 9190-0025
Art. No. 7001-0359

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FD-S 115 and is the answer not in the manual?

Questions and answers

Summary of Contents for Binder FD-S 115

  • Page 1  Fax: +49 7462 2005 100  Internet: http://www.binder-world.com  E-mail: info@binder-world.com  Service Hotline: +49 7462 2005 555  Service Fax: +49 7462 2005 93 555  Service E-Mail: service@binder-world.com  Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3 ...
  • Page 2: Table Of Contents

    Content SAFETY ........................4 Legal considerations ........................... 4 Structure of the safety instructions ...................... 4 1.2.1 Signal word panel ........................4 1.2.2 Safety alert symbol ........................5 1.2.3 Pictograms ..........................5 1.2.4 Word message panel structure ....................6 Localization / position of safety labels on the chamber ..............6 Type plate............................
  • Page 3 13.2.1 Cleaning ..........................25 13.2.2 Decontamination ........................26 13.3 Sending the chamber back to BINDER GmbH ................. 27 14. DISPOSAL......................28 14.1 Disposal of the transport packing ...................... 28 14.2 Decommissioning ..........................28 14.3 Disposal of the chamber in the Federal Republic of Germany ............28 14.4 Disposal of the chamber in the member states of the EU except for the Federal Republic of...
  • Page 4: Safety

    Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall BINDER be held liable for any dam- ages, direct or incidental arising out of or related to the use of this manual.
  • Page 5: Safety Alert Symbol

    WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irreversible) injury CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (reversible) injury CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in damage to the product and/or its functions or of a property in its proximity.
  • Page 6: Word Message Panel Structure

    Figure 1: Position of labels on the chamber front (examples) Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER Service for these replacements. BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019...
  • Page 7: Type Plate

    Made in Germany www.binder-world.com Figure 2: Type plate (example BD-S 115-230V regular chamber) Indications of the type plate (example) Information BINDER Manufacturer: BINDER GmbH BD-S 115 Model designation Incubator Chamber name: Incubator Drying and heating oven Chamber name: Drying and heating oven Serial No.
  • Page 8: General Safety Instructions On Installing And Operating The Chambers

    213-850 on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120 or ZH 1/119, issued by the employers’ liability insurance association) (for Germany). BINDER GmbH is only responsible for the safety features of the chamber provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relat- ing to chamber safety are replaced in the event of failure with original spare parts.
  • Page 9: Intended Use

    Observing the instructions in this operating manual and conducting regular maintenance work (chap. 13) is part of the intended use. WARNING: If customer should use a BINDER chamber running in non-supervised continu- ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two chambers, if this is feasible.
  • Page 10: Chamber Description

    For this, the intended use of the chamber and all its connections must be observed. Chamber description BINDER incubators BD-S / BD-S-UL and drying and heating ovens ED-S / ED-S-UL and FD-S / FD-S-UL are equipped with an electronic PID-controller with digital display.
  • Page 11: Chamber Overview

    Chamber overview (3a) (2a) Figure 3: Chamber overview, closed chamber (4a) (2a) (5a) Figure 4: Chamber overview, open chamber with glass door (BD-S / BD-S-UL) Triangular instrument panel with R-S controller Outer door (2a) Outer door handle Exhaust air outlet (3a) Exhaust air flap Glass door (BD-S / BD-S-UL)
  • Page 12: Triangular Instrument Panel

    Triangular instrument panel Figure 5: Triangular instrument panel with R-S controller Controller display Functional controller buttons Completeness of delivery, transportation, storage, and installa- tion Unpacking, and checking equipment and completeness of delivery After unpacking, please check the chamber and its optional accessories, if any, based on the delivery receipt for completeness and for transportation damage.
  • Page 13: Guidelines For Safe Lifting And Transportation

    Note on second-hand chambers (Ex-Demo-Units): Second-hand chambers are chambers that have been used for a short time for tests or exhibitions. They are thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly. Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before commissioning the chamber.
  • Page 14: Installation

    • Permissible ambient temperature range during operation: +18 °C up to +40 °C / 64.4 °F to 104 °F. At elevated ambient temperature values, fluctuations in temperature can occur. The ambient temperature should not be substantially higher than the indicated ambient temperature of +22 °C +/- 3 °C / 71.6 °F ±...
  • Page 15: Connection To An Exhaust/Ventilation System (Optional)

    Model Model version the power the indicated power frequen- type ber fuse cable 230 V at 50 Hz BD-S 56 BD-S056-230V Grounded plug 6,3 A 230 V at 60 Hz 230 V at 50 Hz ED-S 56 ED-S056-230V Grounded plug...
  • Page 16: R-S Controller Overview

    VDE directives (for Germany). We recommend the use of a residual current circuit breaker. • Only use original connection cables from BINDER. • FD-S / FD-S-UL: Do not place the power cable over the door gap when the chamber is hot after op- eration.
  • Page 17: Menu Structure Overview

    Status-LEDs for information about chamber conditions Heating active Safety controller active Collective alarm Return to Normal display: If no entry is made for 120 seconds, the controller returns from each menu to normal display. To directly return from each menu to normal display, keep pressed down the confirmation button and press the arrow-up button, if appropriate, several times.
  • Page 18: Start Up

    Start up Insert the plug into a suitable socket (chap. 4.2). If there is no indication on the controller, press the confirmation button until the display lights up. The controller now shows normal display (chap. 5). If a ramp gradient was active prior to turning off the chamber, the effective ramp set-point is displayed alternatingly with the target set-point (chap.
  • Page 19: Selecting The Temperature Unit

    If the cut-off temperature is reached, the over temperature protective device permanently turns off the unit. The user cannot restart the device again. The protective cut-off device is located internally. Only a service specialist can replace it. Therefore, please contact an authorized service provider or BINDER service.
  • Page 20: Safety Controller

    Safety controller The chambers are regularly equipped with an adjustable electronic safety controller. It serves to protect the chamber, its environment and the contents against exceeding the maximum permissible temperature. Please observe the DGUV guidelines 213-850 on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120 or ZH 1/119, issued by the employers’...
  • Page 21: Alarm Message And Proceeding In Case Of An Alarm

    9.2.2 Alarm message and proceeding in case of an alarm The status LED “Heating active” is lit. • BD-S / BD-S-UL: Safety controller class 3.1 The safety controller keeps control of the chamber until the chamber temperature cools down below the entered safety controller value.
  • Page 22: Ramp Function

    Ramp function You can program temperature ramps in order to extend heating up times. This may be necessary in some cases to prevent temperature stress in the material during the heating up phase. Temperature ramps should only be used if required. Using them may result in considerably slowing down the heating up times.
  • Page 23: Setting The Gradient

    11.2 Setting the gradient Example: Simultaneously press the confirmation button and the arrow-down button to  access the “Advanced configuration” menu. The password entry display “A.Loc” flashes. Enter the password “20” with the arrow-up button and confirm with the confirma- tion button.
  • Page 24: Options

    Options 12.1 Data logger kits (option) BINDER Data Logger Kits offer an independent long-term measuring system for temperature. They are equipped with a keyboard and a large LCD display, alarm functions and a real-time function. Measure- ment data are recorded in the Data Logger and can be read out after the measurement via the RS232 interface of the Data Logger.
  • Page 25: Cleaning And Decontamination

    We recommend using the neutral cleaning agent Art. No. 1002-0016 for a thorough cleaning. Any corrosive damage that may arise following use of other cleaning agents is excluded from liability by BINDER GmbH. Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER GmbH. CAUTION Danger of corrosion.
  • Page 26: Decontamination

    For chemical disinfection, we recommend using the disinfectant spray Art. No. 1002-0022. Any corrosive damage that may arise following use of other disinfectants is excluded from liability by BINDER GmbH. With every decontamination method, always use adequate personal safety controls.
  • Page 27: Sending The Chamber Back To Binder Gmbh

    13.3 Sending the chamber back to BINDER GmbH If you return a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number (RMA number) that has previously been issued to you. An authorization number will be issued after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us.
  • Page 28: Disposal

    (Elektro- und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739) or contact BINDER service who will organize taking back and disposal of the chamber according to the German national law for electrical and electronic equipment (Elektro- und El- ektronikgerätegesetz, ElektroG from 20 October 2015, BGBl.
  • Page 29: Disposal Of The Chamber In The Member States Of The Eu Except For The Federal Republic Of Germany

    Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739).  Instruct BINDER Service to dispose of the device. The general terms of payment and delivery of BINDER GmbH apply, which were valid at the time of purchasing the cham- ber.
  • Page 30: Disposal Of The Chamber In Non-Member States Of The Eu

    14.5 Disposal of the chamber in non-member states of the EU CAUTION Alteration of the environment.  For final decommissioning and disposal of the chamber, please contact BINDER Ser- vice.  Follow the statutory regulations for appropriate, environmentally friendly disposal.
  • Page 31: Troubleshooting

    “OPEN” on upper Contact BINDER service. sor and controller. display. Only qualified service personnel authorized by BINDER must perform repair. Repaired chambers must comply with the BINDER quality standards. BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 31/59...
  • Page 32: Technical Description

    (certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 sys- tems. They are controlled and calibrated to a DKD-Standard at regular intervals.
  • Page 33: Technical Data

    16.4 Technical data Model BD-S ED-S FD-S BD-S ED-S FD-S BD-S-UL ED-S-UL FD-S-UL BD-S-UL ED-S-UL FD-S-UL Chamber size Exterior dimensions mm / 595 / 595 / 595 / 745 / 745 / 745 / Width, net inch 23.43 23.43 23.43 29.33 29.33 29.33...
  • Page 34 Model BD-S ED-S FD-S BD-S ED-S FD-S BD-S-UL ED-S-UL FD-S-UL BD-S-UL ED-S-UL FD-S-UL Chamber size Temperature data (continued) Temperature uniformity (varia- +/- K tion) at 37 °C / 98.6 °F Temperature uniformity (varia- +/- K tion) at 150 °C / 302 °F Heating up time to 150 °C / minutes 302 °F...
  • Page 35: Equipment And Options (Extract)

    If the chamber is fully loaded, the specified heating up times may vary according to the load. 16.5 Equipment and options (extract) To operate the chamber, use only original BINDER accessories or accessories / components from third-party suppliers authorized by BINDER. The user is responsible for any risk arising from using unauthorized accessories. Standard equipment R-S microprocessor temperature controller Timer function “Delayed Off”...
  • Page 36: Dimensioned Drawings

    16.7 Dimensioned drawings 16.7.1 BD-S / BD-S-UL 56 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 36/59...
  • Page 37: Bd-S / Bd-S-Ul 115

    16.7.2 BD-S / BD-S-UL 115 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 37/59...
  • Page 38: Ed-S / Ed-S-Ul 56

    16.7.3 ED-S / ED-S-UL 56 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 38/59...
  • Page 39: Ed-S / Ed-S-Ul 115

    16.7.4 ED-S / ED-S-UL 115 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 39/59...
  • Page 40: Fd-S / Fd-S-Ul 56

    16.7.5 FD-S / FD-S-UL 56 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 40/59...
  • Page 41: Fd-S / Fd-S-Ul 115

    16.7.6 FD-S / FD-S-UL 115 [mm] BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 41/59...
  • Page 42: Certificates And Declarations Of Conformity

    Certificates and declarations of conformity 17.1 EU Declaration of Conformity for BD-S BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 42/59...
  • Page 43 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 43/59...
  • Page 44: Eu Declaration Of Conformity For Ed-S

    17.2 EU Declaration of Conformity for ED-S BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 44/59...
  • Page 45 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 45/59...
  • Page 46: Eu Declaration Of Conformity For Fd-S

    17.3 EU Declaration of Conformity for FD-S BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 46/59...
  • Page 47 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 47/59...
  • Page 48: Certificate For The Gs Mark Of Conformity Of The "Vde Prüf- Und Zertifizierungsinstitut" (Testing And Certification Institute Of The Association For Electrical, Electronic And Information Technologies)

    17.4 Certificate for the GS mark of conformity of the “VDE Prüf- und Zertifizier- ungsinstitut” (Testing and Certification Institute of the Association for Elec- trical, Electronic and Information Technologies) BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 48/59...
  • Page 49 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 49/59...
  • Page 50 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 50/59...
  • Page 51 BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 51/59...
  • Page 52: Certificate Of Ul Compliance From Underwriters Laboratories Inc

    ® 17.5 Certificate of UL Compliance from Underwriters Laboratories Inc. Valid for UL chambers only BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 52/59...
  • Page 53: Product Registration

    Product registration BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 53/59...
  • Page 54: Contamination Clearance Certificate

    Contamination clearance certificate 19.1 For chambers located outside the USA and Canada Declaration with regard to safety and health Erklärung zur Sicherheit und gesundheitlichen Unbedenklichkeit The German Ordinance on Hazardous Substances (GefStofV), and the regulations regarding safety at the workplace, require that this form be filled out for all products that are returned to us, so that the safety and health of our employees can be warranted.
  • Page 55 Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.) ___________________________________________________________________________________ Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH: ___________________________________________________________________________________ BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019...
  • Page 56 We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been entrusted with the handling / repair of the chamber / component. / Es ist uns bekannt, dass wir gegenüber...
  • Page 57: For Chambers Located In The Usa And Canada

    Please complete this form and the Customer Decontamination Declaration (next 2 pages) and attach the required pictures. E-mail to: IDL_SalesOrderProcessing_USA@binder-world.com After we have received and reviewed the complete information we will decide on the issue of a RMA number. Please be aware that size specifications, voltage specifications as well as performance specifi- cations are available on the internet at www.binder-world.us...
  • Page 58 Customer (End User) Decontamination Declaration Health and Hazard Safety declaration To protect the health of our employees and the safety at the workplace, we require that this form is com- pleted by the user for all products and parts that are returned to us. (Distributors or Service Organizations cannot sign this form) NO RMA number will be issued without a completed form.
  • Page 59 4.5 Shipping laws and regulations have not been violated. I hereby commit and guarantee that we will indemnify BINDER Inc. for all damages that are a consequence of incomplete or incorrect information provided by us, and that we will indemnify and hold harmless BINDER Inc.

Table of Contents