Operating Instructions
12 Appendices
12.1
Appendix 1 – Jumper Positions
Jumper Positions - Internal to the controller
JP 1
JP 2
JP 3
Fuse
68X216802
Selects the input voltage 220 VAC.
Selects the input voltage 110 VAC.
Selects the temperature sensor for Pt1000/Pt100
Note that there is a fuse (slow-blow 100mA) internal to the controller. Before
opening the unit, ENSURE that the power cable is physically separated from
the power supply. Replace fuse with the recommended type only.
lpha-CON1000