Operating Instructions
7.4

Controller (Cntr) sub-function

You can set the controller's parameters in this sub-function.
7.4.1
Entering the Controller sub-function
Enter Advanced Set-up mode. Push the ENTER key and scroll to Advanced set-up security code
1)
"22". Push the ENTER key again.
2)
Press the
or
7.4.2
Choosing the controller type (limit or monitor)
This mode lets you choose your controller type: limit control or control off.
Use control off to operate controller as a monitor only or to prevent relays from switching.
-
1) Follow directions in 7.6.1 to enter Controller mode.
2) Press the ENTER key. The upper display shows the current controller type and the lower display
shows "tyP".
3) Press the
or
L.Ct = limit value pickup (on/off control).
-
oFF = controller off.
-
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.6.3, or return to Measurement mode by pressing the
(escape).
68X216802
keys to scroll until the upper display shows "Cntr".
HOLD SETUP
Cntr
Cntr
keys to select your controller type.
HOLD SETUP
L.Ct
L.Ct
tyP
tyP
HOLD SETUP
oFF
oFF
tyP
tyP
HOLD SETUP
dEEn
dEEn
rEL
rEL
and
keys simultaneously
lpha-CON1000
19