ADVANCED SET-UP MODE

7.1

Temperature Coefficient sub-function

This sub-function allows you to select the correct temperature coefficient for optimum operations. For
applications in the pure water or ultr-pure water industries, simply select the "Pur" temperature
coefficient option. For all other applications, select "Lin" temperature coefficient. The controller allows
further input of temperature coefficient values, independently for the process and for the calibrating
solutions. Default is at 2.10 %.
HOLD SETUP
tc tc
7.1.1
Selecting Pure-water or Linear Temperature Coefficient
HOLD SETUP
HOLD SETUP
HOLD SETUP
HOLD SETUP
68X216802
HOLD SETUP
Lin
Lin
tC tC
HOLD SETUP
Pur
Pur
tC tC
1) Enter Advanced set-up mode. Push the ENTER key and
tc tc
scroll to Advanced Set-up security code "22". Push the ENTER
key again. Controller displays "Lin" in the upper display and "tC"
in the lower display.
2) Press the
temperature coefficients. If "Pur" tC is be selected, press the
ENTER key and controller goes back to "tC". If "Lin" tC is selcted,
Lin Lin
press the ENTER key and proceed to step 3.
3) The upper display shows "2.10%", while the lower display
tC tC
show "P.tC". This option allows you to input the Process
Temperature coefficient, from 0 to 10%. Default is 2.10%. Use the
or
4) Press the ENTER key. Next, enter the calibration solution
2.10
2.10
temperature coefficient value, from 0 to 10%. Default is 2.10%.
% %
5) Press the
P.tC
P.tC
solution temperature coefficient.
ATC
ATC
6) Press the ENTER key to accept the value.
7)
to Measurement mode by pressing the
simultaneously.
2.10
2.10
% %
C.tC
C.tC
ATC
ATC
HOLD SETUP
HOLD SETUP
2.10
2.10
%
P.tC
P.tC
ATC
or
keys to enter the required temperature coefficient value.
or
Continue with additional Advance setup procedures or return
2.10
2.10
%
C.tC
C.tC
ATC
keys to select between "Pur" and "Lin"
keys to input the desired calibration
and
keys (escape)