Operating Instructions
11.4 Guidelines for Returning Unit for Repair
Use the original packaging material, if possible when shipping the unit for repair. Otherwise wrap it
with bubble pack and use a corrugated box for better protection. Include a brief description of any
faults suspected for the convenience of Customer Service Dept., if possible.
12 APPENDICES
ENSURE that the power cable is physically separated from the power
.
supply
12.1 Appendix 1 – Jumper Positions
Jumper Positions - Internal to the controller
JP 1
JP 2
JP 6
Fuse
JP6
On the Analog PCB:
Selects between Pt100 and Pt1000. To switch between Pt100 and Pt1000, use
tweezers or needle nose pliers to shift jumper from Position A to Position B.
JP2 JP1
Fuse
JP6
B
A
View from the front of the Analog PCB
Selects the input voltage 220 VAC.
Selects the input voltage 110 VAC.
Selects the temperature sensor for Pt1000/Pt100
Note that there is a fuse (slow-blow 100mA) internal to the controller. Replace
fuse with the recommended type only.
View from the top of Main PCB
Analog PCB
JP6
A
- Pt100
B
- Pt1000
28