Calibration (Cal) Sub-Function - EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER Instruction Manual

Conductivity controller/transmitter
Hide thumbs Also See for ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER:
Table of Contents

Advertisement

Operating Instructions
HOLD SETUP
Stdy
Stdy
AL.C
AL.C
HOLD SETUP
FLEt
FLEt
AL.C
AL.C
7.7.4
Input Line Resistance Adjust
This function compensates for the line resistance of the cable to its cell.
HOLD SETUP
0.0 0.0
L.Ad
L.Ad
5)
Proceed to 7.9.5, or return to Measurement mode by pressing the
(escape).
7.7.5
Reverting to factory default settings
Use this parameter to reset all settings to factory default. Changing from "no" to "YES" and pressing the
ENTER key resets all settings to factory default.
HOLD SETUP
dEF
dEF
3)
Press the
or
overwrite all setting you have made and automatically return you to Measurement mode.
4)
Press the ENTER key to confirm your selection and to return to Advanced Set-up mode, or return
to Measurement mode by pressing the
7.8

Calibration (CAL) sub-function

The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration
mode. The only difference is that the controller will revert back to Set-up mode (instead of Measurement
mode) after calibration is completed.
68X216802
1)
Follow directions in 7.9.2 to enter Configuration mode.
2)
Press the ENTER key. Scroll with the
display shows "Stdy" or "FLEt" and lower display shows "AL.C.".
AL.C = alarm contact
-
StdY = steady contact
-
FLEt = fleeting (single pulse) contact
-
Press the
3)
4)
Press the ENTER key to confirm your selection.
5)
Proceed to 7.9.4, or return to Measurement mode by pressing the
and
1)
Follow directions in 7.9.3 to enter Configuration mode.
Press the ENTER key. Scroll with the
2)
upper display shows "0.0" and lower display shows "L.Ad".
L.Ad = line adjuster resistance (0.0 to 100.0)
-
3)
Press the
Press the ENTER key to confirm your selection.
4)
WARNING: If you select yes, all the settings you have made will be
overwritten as a result!
no no
Follow directions in 7.9.4 to enter Configuration mode.
1)
2)
Press the ENTER key. Scroll with the
display shows "no" or "YES", and the lower display shows "deF"
(default).
keys to select no or yes. Selecting "YES" and pressing the ENTER key will
or
keys to select steady or pulse contact.
keys simultaneously (escape).
or
keys to input the value.
and
keys simultaneously (escape).
lpha-CON1000
or
keys until the upper
or
keys until the
and
keys simultaneously
or
keys until the upper
23