Measurement Range Sub-Function - EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER Instruction Manual

Conductivity controller/transmitter
Hide thumbs Also See for ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER:
Table of Contents

Advertisement

7.4.3
Choosing break/make contact relay type
Note: If the controller type "oFF" is set, the parameters listed in 7.5.3, 7.5.4, 7.5.5 and 7.5.6 are blanked
out.
This mode lets you determine the relay-state under Non-Alarm condition – dEEN (de-energised) or EN
(energised).
HOLD SETUP
dEEn
dEEn
rEL
rEL
HOLD SETUP
En En
rEL
rEL
7.5

Measurement Range sub-function

In this sub-function, the appropriate range is selected with the appropriate cell constant.
HOLD SETUP
rng
rng
HOLD SETUP
200
200. . 0 0
0.1 0.1
4
Proceed to 7.8 to set current output or
5)
keys simultaneously (escape).
68X216802
1) Follow directions in 7.6.2 to enter Controller mode.
2) Press the ENTER key. Scroll until the lower display shows "rEL"
and the upper display shows the current selection (de-energised =
dEEN or energised = EN).
3) Press the
relay state.
4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to
Measurement mode by pressing the
(escape).
7.5.1
Selecting the Measuring Range sub-function
Enter Advanced Set-up mode. Push the ENTER key and scroll to
1)
Advanced Set-up security code "22". Push the ENTER key again.
2)
Press the
"rng".
3)
Press the ENTER key. The upper display shows the maximum
measurement range, while the lower display shows the Cell constant. In
S
the lower left-corner of the LCD, the number corresponding to the
respective range is displayed.
Press the
4)
Appendix 2 for the full list of measurement ranges).
return to Measurement mode by pressing the
or
keys to choose de-energised or energised
or
keys to scroll until the upper display shows
or
keys to select the correct range. (Please refer to
and
keys simultaneously
and

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alpha-con1000

Table of Contents