Measurement Range sub-function

In this sub-function, the appropriate range is selected with
the appropriate cell constant.
7.5.1

Selecting the Measuring Range sub-function

1)
Enter Advanced Set-up mode. Push the ENTER key and scroll
to Advanced Set-up security code "22". Push the ENTER key
again.
2)
Press the ▲ or ▼ keys to scroll until the upper display shows
"rng".
3)
Press the ENTER key. The upper display shows the maximum
measurement range, while the lower display shows the Cell
constant. In the lower left-corner of the LCD, the number
corresponding to the respective range is displayed.
4)
Press the ▲ or ▼ keys to select the correct range. (Please refer to Section 12.2 Appendix 2 for
the full list of measurement ranges).
5)
Proceed to 7.6 to set current output or return to Measurement mode by pressing the ▲ and ▼
keys simultaneously (escape).