Shodex PROTEIN LS-G 4J Operation Manual

Advertisement

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.
1. Introduction
Thank you for purchasing the Shodex product. Shodex PROTEIN LW-403 4D is a silica-based gel filtration
chromatography column. The column is packed with small particles having carefully optimized pore-sizes. It is
suitable for rapid analysis of proteins with a molecular weight of several hundred of thousand.
2. Column Components
Please refer to the Shodex website: https://www.shodex.com/en/da/07.html
3. Column Specifications
Product
Product Name
Code
PROTEIN LW-403 4D
F6989403
PROTEIN LS-G 4J
F6700134
(Guard Column)
Base Material
: Spherical porous particles of silica modified with hydrophilic group
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water
4. Usable Conditions
Product Name
PROTEIN LW-403 4D
PROTEIN LS-G 4J
Operation Manual
Shodex PROTEIN LW-403 4D
Column Size
Particle
(mm)
Size
(μm)
I.D.
Length
4.6
150
1.9
4.6
20
1.9
Flow Rate (mL/min)
Recommended
Maximum
0.35
-
Resonac Corporation (https://www.shodex.com/)
Theoretical
Target Molecular
Plate Number
Weight Range
(Per Column)
≥ 11,000
≥ 1,300
*1
Reference value only / Measured with proteins
Maximum
Pressure
(MPa/column)
0.5
20
-
-
ver.20221202E
Exclusion
*1
Limit
10,000 - 700,000
(1,000,000)
-
*2
Estimated value
pH
Temperature
Range
Range (°C)
3.0 - 7.5
4 - 45
*1
*2
-
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROTEIN LS-G 4J and is the answer not in the manual?

Questions and answers

Summary of Contents for Shodex PROTEIN LS-G 4J

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex PROTEIN LW-403 4D is a silica-based gel filtration chromatography column. The column is packed with small particles having carefully optimized pore-sizes. It is suitable for rapid analysis of proteins with a molecular weight of several hundred of thousand.
  • Page 2: Eluent Preparation

    This may lead to abnormal peak shapes, insufficient separations, and/or low repeatability. To avoid the negative influences, it is recommended to lower the sample concentration or alternatively, lower the injection volume. Resonac Corporation (https://www.shodex.com/)
  • Page 3: Column Installation

    Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. Resonac Corporation (https://www.shodex.com/)
  • Page 4: Column Cleaning

    Securely tighten the end caps and store the column at a location with stable temperature (a cool and dark space is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. Attention! · Never allow inside the column to dry. It can damage the column. Resonac Corporation (https://www.shodex.com/)
  • Page 5 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...

This manual is also suitable for:

Protein lw-403 4dF6989403F6700134

Table of Contents