Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex HILICpak VG-50 series is a polymer-based HILIC column modified with amino group. The column is suitable for the analysis of saccharides and sugar alcohols.
Page 2
(4) When a sample contains protein or lipid, make sure to remove them prior to the injection. Presence of those substances may damage the column. Note · Use of guard column is recommended to protect the analytical column. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 3
This may result in creating an empty space in the column and deteriorates the column. Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
VG-50 4E 6 min 10 min 0.1 M HNO aq. solution 75 min 120 min 6 min 10 min 0.1 M NaOH aq. solution 75 min 120 min 12 min 20 min Eluent 40 min 60 min SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.