Operation Manual Shodex RSpak NN-814 Version 211102E Operation Manual Shodex™ RSpak™ NN-814 Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
Page 2
Certificate of Analysis (CoA) are matching and correct. Introduction Thank you for purchasing the Shodex™ product. Shodex™ RSpak™ NN-814 is a multimode (reversed phase and cation exchange modes) column filled with spherical porous particles of polyhydroxymethacrylate modified with sulfo group. It is an ideal column for analyzing complex samples that contain mixture of neutral and ionic substances.
Page 3
Operation Manual Shodex RSpak NN-814 Version 211102E Usable conditions Temperature Flow Rate (mL/min) Maximum Pressure Product Name Range Range (°C) (MPa) (Per Column) Recommended Maximum 2 - 12 10 - 60 RSpak NN-814 0.5 - 1.0 2 - 12 10 - 60 RSpak NN-G Usable solvents are listed below.
Page 4
Operation Manual Shodex RSpak NN-814 Version 211102E (2) Filter the sample solution using disposable 0.45 μm filter to prevent the column from clogging or deteriorating. (3) Recommended sample injection volume is 50 μL per column. Note • Use of guard column is recommended to protect the analytical column.
Page 5
Operation Manual Shodex RSpak NN-814 Version 211102E temperature return to room temperature before stopping the pump. If the pump was stopped while the eluent inside the column was still hot, as the eluent temperature decreases, its volume also decreases. This may result in creating an empty space in the column and deteriorates the column.
Page 6
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex website (www.shodex.de) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact...
Page 7
11. Warranty (Version 4) (1) Showa Denko K.K. warrants that the Shodex™ product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.