Page 1
Operation Manual Shodex SUGAR EP SC1011-7F Version 210415E Operation Manual Shodex™ SUGAR EP SC1011-7F Sales Office: Technical Support: Showa Denko Europe GmbH info@shodex.de Shodex Business +49 (0)89 / 93 99 62 37 Konrad-Zuse-Platz 3 81829 Munich Quotes and Orders: Europe, Middle East, Africa, Russia Germany order@shodex.de...
Page 2
Certificate of Analysis (CoA) are matching and correct. Introduction Thank you for purchasing the Shodex™ product. Shodex SUGAR EP SC1011-7F is a high performance chromatography column which is specifically designed for the separations and analysis of sugars. A strong acid cation exchange type packing gel, prepared from a rigid resin of styrene divinylbenzene copolymer provides a mixed separation mode.
Page 3
Operation Manual Shodex SUGAR EP SC1011-7F Version 210415E Base Material: Spherical porous particles of styrene divinylbenzene copolymer modified with sulfo group (counter ion: Ca Column Housing: SUS-316 (stainless steel) Screw Type: Internally-threaded type No. 10-32 UNF Shipping Solvent: Water Usable Conditions Flow Rate (mL/min) Temperature (°C)
Page 4
Operation Manual Shodex SUGAR EP SC1011-7F Version 210415E Sample Preparation (1) If possible. use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes the sample.
Page 5
Operation Manual Shodex SUGAR EP SC1011-7F Version 210415E Presence of an extra space will let the sample to spread out and may result in wide peaks. (3) Set the initial flow rate at 0.2 - 0.3 mL/min and start the system. If using the column at an elevated temperature, keep a low flow rate until the temperature of the column reaches to the set temperature, and then gradually increase the flow rate to the desired.
Page 6
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex website (www.shodex.de) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact...
Page 7
3-3 If the Shodex product is disposed of 3-4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K.K. 3-5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below: a.