BEYES Maxso E800P Instructions For Use Manual

Portable electric handpiece system

Advertisement

Quick Links

Maxso
E800P
®
Portable Electric Handpiece System
Instruction for Use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Maxso E800P and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BEYES Maxso E800P

  • Page 1 Maxso E800P ® Portable Electric Handpiece System Instruction for Use...
  • Page 2: Table Of Contents

    7 Cleaning, disinfection and sterilization ........14 8 Troubleshooting ................15 9 Storage and transport ..............15 10 After-sales service ..............15 11 Environment protection ............. 16 12 Symbol instruction ..............17 13 EMC-Declaration of comformity..........17 14 Beyes Limited Warranty Statement .......... 22...
  • Page 3: Introduction

    2. Maxso E800P requires special precautions for EMC and needs to be installed and put into use according to the EMC environment.
  • Page 4 1.3 Model Maxso E800P 1.4 Contraindications Beyes Dental Canada Inc. will not be liable for any direct or indirect damages and losses under the following conditions: •The device is used to the unmentioned usages or the usages outside the scope of applications.
  • Page 5: Basic Technical Parameters

    2. Basic technical parameters 2.1 Specification of main unit Model: Maxso E800P Size:165.5mm×129.7mm×77.6mm Power supply input: 100-240V~ 50/60Hz 2.5A Power supply output: DC 30V 3.2A 2.2 Motor specification Model: E800 MC Rotation speed: 2000-40000 rpm; Voltage input: DC 24V Size: φ22×76.7mm Tail cord length: 1.8m...
  • Page 6 3.2 See the packing list for the machine configuration. It mainly consists of main unit, motor, power adapter, etc. Motor tail Motor Silicone handle Main unit Contra-angle bracket handpiece Power adapter Figure 1 3.3 Front View of the Main Unit Motor tail Motor Contra-angle...
  • Page 7 3.4 Rear view of the main unit Motor tail Four-hole DC adapter Power switch line interface handpiece connector interface Figure 3 3.5 Schematic diagram of motor installation Motor Cord Motor Cord Motor Connector Figure 4-1 Motor Cord 3.5.1 Connecting/ Disconnecting the motor and the motor cord A.
  • Page 8: Main Unit Interface

    3.6 Installation steps 3.6.1 First open the package to check whether the items of the equipment are complete according to the packing list, and place the main unit on a stable surface. 3.6.2 Connect the four-hole handpiece tube to the control box through the four- hole handpiece connector and tighten it.
  • Page 9 4.3 Manual mode interface DEMO MODE Figure 7 4.4 Interface of calibration mode Please click "start" after fully stepping on the foot pedal for calibration. St ar t exit Figure 8 Figure 9 Enter the setting interface, click the calibration mode touch button, enter the air pressure calibration interface, click “Start”, and the interface will pop up as shown in Figure 9.
  • Page 10: Function And Operation

    4.5 Restore the factory setting Are you sure you want to restore to factory setting? Figure 10 Enter the setting interface, and click the restore factory settings touch button to enter the restore factory settings interface for confirmation. In confirming interface, click OK, the interface shown in Figure 10 will pop up.
  • Page 11 5.3 The electric motor operation is controlled by the foot pedal of the dental chair. 5.4 Make sure that the pedal control calibration is performed before using the unit for the first time. 5.5 Icon description Icon Name Function Mode Select the preset fixed speed.
  • Page 12 5.6 Basic function adjustment on main unit controlling interface 1. Speed ratio 2. Speed ratio 7. Setting 6. Store 3. Speed adjustment 5. Mode 4. Forward 5.6.1 Select speed ratio by through touching key “1”. The color of the key will change when touching the key.
  • Page 13: Safety Precautions

    6.2 The accuracy of the speed monitoring depends on the high precision performance of the handpiece installed on the micro motor. To ensure the accuracy of the display speed, please use Beyes electric handpieces with the motor. 6.3 Read this operating manual before use and fully understand the functions of each part.
  • Page 14: Cleaning, Disinfection And Sterilization

    7. Cleaning, disinfection and sterilization 7.1 Cleaning Cleaning the motor surface with a water-absorbent cloth to remove any liquid residue from the motor surface and wipe it back and forth 5 times. 7.2 Disinfection Use a chlorine-free disinfectant to wipe the motor. It is recommended to use Cavicide disinfectant or 75% alcohol, but do not soak the motor in the disinfectant.
  • Page 15: Troubleshooting

    9.6 Avoid being exposed to sun, rain, and snow during transport. 10. After-sales service Beyes Dental Canada Inc. is responsible for repairs under the warranty period if the device cannot work normally due to a quality defect. The device must be registered with Warranty Card from the date of invoice to the original user.
  • Page 16: Environment Protection

    11. Environment protection Part Toxic or harmful substances or elements (Pb) (Hg) (Cd) (Cr6+) (PBB) (PBDE) ○ ○ ○ ○ ○ ○ Main unit ○ ○ ○ ○ ○ ○ Motor handpiece ○ ○ ○ ○ ○ ○ Power adapter ○...
  • Page 17: Symbol Instruction

    Table 1: Declaration - electromagnetic emissions Guidance and manufacturer’s declaration - electromagnetic emissions The model Maxso E800P is intended for use in the electromagnetic environment specified below. The customer or the user of the model Maxso E800P should assure that it is used in such an environment.
  • Page 18 Table 2: Guidance & Declaration - electromagnetic immunity Guidance & Declaration — electromagnetic immunity The model Maxso E800P is intended for use in the electromagnetic environment specified below. The customer or the user of the model Maxso E800P should assure that It is used in such an environment.
  • Page 19 Conducted RF & Radiated RF Guidance & Declaration - Electromagnetic immunity The model Maxso E800P is intended for use in.the electromagnetic environment specified below. The customer or the user of the models Maxso E800P should assure that it is used in such an environment.
  • Page 20 Immunity test IEC 60601 Compliance Electromagnetic test level level environment - guidance Conducted RF 3 Vrms Portable and mobile RF lEC 61000-4-6 150 kHz to communications equipment Conducted RF 80 MHz 6 3V/m should be used no closer to lEC 61000-4-6 Vrms ISM any part of the models Maxso Radiated RF...
  • Page 21 E800P should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model Maxso E800P. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
  • Page 22: Beyes Limited Warranty Statement

    25 months. 14.3 EXCLUSIONS This limited warranty does not cover and BEYES shall not be liable for the following; 1. Defects, damage or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, negligent storage, tampering or failure to seek and obtain repair or replacement in a timely manner;...
  • Page 23 12. Products that would otherwise by covered under Sections 1 and 2 of this limited warranty, but are acquired: (i) from a person or entity that is not BEYES or one of its authorized dealers; or (ii) from a BEYES dealer that is not authorized...
  • Page 24 Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. Lotus NL B.V.