Page 26
(inch) (lb) mm (inch) kg (lb) Hydraulic Power 670 mm (26½”) 670 mm (26½”) Overall width 710 mm (28”) 710 mm (28”) Length 1130–1350 mm (44½–53¼”) 1080–1350 mm (44½–53¼”) Overall height 550 mm (21¾”) 550 mm (21¾”) Effective seat width Seat width between hip supports 350–490 mm (13¾...
Page 29
Applicable requirements examples: Presales information and instructions • Material safety (biocompatibility, flammabil- Presales information and instructions are avail- ity), mechanical safety (stability, squeezing able at www.etac.com, under your product, in hazards, inter connection), electrical safety. the documents section. Disposal When a product reaches the end of its lifetime,...
Page 30
Safety Signs, symbols and instructions placed on this product must never be covered or removed and must remain present and clearly legible throughout the entire lifetime of the product. Immediately replace or repair illegible or damaged signs, symbols and instructions. Contact your local dealer for instructions. Any adverse event in relation to the device must be reported to your local dealer and the national com- petent authority in a timely manner.
Service & maintenance information Service and maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to service and maintain a device in accordance with the instructions may invalidate the device’s warranty. Furthermore, failure to service and maintain a device may compromise the clinical condition or safety of users and/or their caregivers.
Power Unit system Handset functionality • Battery is OK, no need for charging. Green LED lights when activating functions • Battery is soon empty, the battery needs recharging. Orange LED Lights when activating functions • The battery needs recharging, red LED flashes when activating func- tions.
Technical Data Ambient and surface temperature The table below illustrates how long time it Battery: BAJ1 - IPX5 takes for the surfaces of the Heron to reach • Capacity 2.9 Ah - 24V DC an ambient temperature of 20°C from -10°C •...
Page 35
Electromagnetic compatibility (EMC) envi- EMC - WARNINGS Use of this equipment adjacent to or stacked ronment with other equipment should be avoided This equipment is for use in home healthcare because it could result in improper operation. If facility environment. such use is necessary, this equipment and the This equipment is not likely to cause any harmful other equipment should be observed to verify...
Page 36
Test modes during immunity tests: Battery mode, standby armed/ready. Charging mode, actuator not running. EM immunity standards IEC 60601-1-2:2014 test Additional test levels levels Electrostatic discharge ± 8 kV contact ± 6 kV contact IEC61000-4-2 ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air Radiated RF EM fields 3V/m (Professional)