Tektronix ADA400A Instruction Manual page 58

Differential preamplifier
Hide thumbs Also See for ADA400A:

Advertisement

CAUTION. To prevent damaging internal components by means of electrostatic
discharge (ESD), perform all service at a static-free work station and wear a
grounded anti-static wrist strap.
Do not slide the internal module over any surface. Place internal module in
static-free bag for transportation away from work station.
cleaners.
Dirt on external surfaces may be removed with a soft cloth dampened with a
mild detergent and water solution, or isopropyl! alcohol.
Alcohol and rinse with warm deionized water.
46
Interface Pins
on page 57.
Periodically check to see that each of the interface pins move freely and fully
should be replaced.
To remove a TEKPROBE interface pin, firmly grasp the pointed tip with pliers
and pull the pin out of the connector. See Figure 17.
No tools are required to install a replacement pin. Insert a new pin into the
pin into the connector by pressing the tip gently but firmly against a hard
surface, such as a wood block or table top.