Provides Low Level Signal
Acquisition Capability to DSOs
10 µV/div Sensitivity
Low Pass Filters for 100 Hz,
3 kHz, 100 kHz, and Full (1 MHz)
CMRR: > 100 dB DC to 10 kHz
Selectable Gain:
100, 10, 1, or 0.1
Accurately Measures Signals
As Small As 5 µV
Ideal for Medical Research,
Mechanical Testing, and
Medical Equipment Design
No Additional Power Supplies or
Cables Required When Used
with Oscilloscopes That Have a
TekProbe
™
BNC Connector
Interface
ow low level differen-
N
tial measurements
can be made on any Tek
scope. The ADA400A pre-
amplifier is perfect for low
frequency differential mea-
surements or low ampli-
tude, single ended mea-
surements with lots of
noise.
The ADA400A preamplifier
is designed for biophysical
research and mechanical
transducer, medical diag-
nostic, and imaging equip-
ment design, as well as
audio equipment design
applications.
F A X a b l e
ADA400A Differential Preamplifier
STANDARD
SPECIFICATIONS
Gain Settings
0.1, 1, 10, 100
Differential Signal Range
(with DC Offset off)
0.1
± 80 V
1
± 10 V
10
± 1 V
100
± 100 mV
Common Mode Signal Range
0.1, 1
± 40 V
10, 100
± 10 V
Power Supply
TekProbe Power
(for non-TekProbe BNC
connector scopes use 1103
Power Supply)
Electrostatic Immunity
(per IEC 801-2)
> 8 kV
DC Gain Accuracy
± 2%
D a t a
S h e e t
TYPICAL
CHARACTERISTICS
CMRR
(at gain of 100)
> 100 dB (100,000:1) at 10 kHz
> 90 dB (30,000:1) at 100 kHz
Input R
1 MΩ (each input)
∞
Ω jumper selectable)
(
Input C
55 pF ± 20% (each input)
Offset Range
0.1, 1
± 40 V
10, 100
± 1 V
Bandwidth
DC to > 1 MHz
Selectable Filters
100 Hz, 3 kHz, 100 kHz
Output Load Requirements
1 MΩ
PHYSICAL
CHARACTERISTICS
Output Cable Length
2 m (6.6 ft)
Weight
kg
lb
Net
1.49
3.3
Dimensions
cm
in
Height
3.6
1.4
Width
11.2
4.4
Length
20.3
8.0
1
Need help?