Performance Verification
Requirements fo
Table 8: Test Equipment
r Performance
The performance limits of this procedure are valid with the following conditions:
a
The ADA4OOA Differential Preamplifier is in an environment that is within
the operating limits of temperature, altitude, humidity, and vibration
=
©The data line jumper located inside the ADA400A Differential Preamplifier.
is properly configured depending on the type of oscilloscope you connect to
to warm up prior to verification.
Differential Preamplifier. The Adjustment Procedures section begins on page 35.
Description
Minimum Requirements
Example
Purpose
Bandwidth: > 5 MHz
TDS460A
Supply power to and display
Sensitivity: 1 mV/div
output of ADA400A
Auto Measurement: Frequency,
input Impedance: 1 MQ
2.
Sine Wave Generator
5 MHz
3.
Digital Multimeter
Resolution: 2 3.5 digits
DM2510
Input resistance check
10MQ
ADA4O0A Instruction Manual
27