Job Aid
BD FACSDiscover S8 Cell Sorter:
Performing a daily system shutdown
This job aid contains instructions for how to perform a daily shutdown. A daily shutdown cleans the sample path and fills the flow cell
®
with BD
Detergent Solution in preparation for shutdown.
NOTE If you are shutting down the system for more than two days, select the long-term shutdown procedure. See the job aid for
performing a long-term shutdown.
For additional information, see the Maintenance section in the BD FACSDiscover S8 Cell Sorter with BD CellView Image
Technology and BD SpectralFX Technology User's Guide.
S t a r t u p
S e t u p
s y s t e m
e x p e r i m e n t
Materials Needed
• 10% bleach (approximately 3 mL)
• Deionized (DI) water
• BD
®
Detergent Solution Concentrate
Before you begin
®
Prepare 3 mL of 1.5% BD
Detergent Solution from the BD
the instructions on the label for correct dilution information and see instead the safety data sheet regarding handling.
®
CAUTION Never mix BD
Detergent Solution and bleach because it can create dangerous fumes.
Cleaning the sample line
This procedure cleans the sample line with a bleach solution. We recommend cleaning the sample line at the end of each experiment,
between users, and before shutting down.
Load a tube containing 3 mL of a 10% bleach solution onto the sample loading port.
1.
From an open experiment, from the View Data tab, click Load.
2.
3.
After approximately 5 minutes, click Unload.
Load a tube containing 3 mL of DI water onto the sample loading port.
4.
NOTE It is important to always run a tube of DI water after running bleach on the cell sorter.
Repeat steps 2 and 3.