Page 4
A.1 About these instructions....................29 A.2 Document conventions ....................29 A.3 Advisory information ......................30 A.4 Unpacking the BD PrepMate™ Automated Accessory ........... 31 A.5 Initial setup and test run ....................32 A.6 Load BD PrepMate Automated Accessory for test run ..........32...
The BD PrepMate™ Automated Accessory mixes and removes the specimen from a BD SurePath™ Collection Vial or BD CytoRich™ Clear Vial. It then layers the specimen onto the density reagent in a centrifuge tube. The automated process handles from one to twelve specimens per cycle.
Page 6
BD PrepMate™ Instrument User’s Manual Hold down Specimen rack door latch Tray latch On/Off switch Figure 1 – BD PrepMate™ Automated Accessory front view Power cord inlet Figure 2 – BD PrepMate™ Automated Accessory rear view...
Accessory. The BD PrepMate™ Automated Accessory is not a stand-alone system; it is a component of the BD SurePath™ Liquid-based Pap Test and may be used with either the BD PrepStain™ Slide Processor or the BD Totalys™ SlidePrep to prepare BD SurePath™ slides. Likewise, this manual is not intended to be a stand-alone document;...
Indicates the possibility of severe equipment damage or invalid results if instructions are not followed. NOTE Gives helpful information about the BD PrepMate™ Automated Accessory. Symbol glossary There are several symbols used on the instrument. Those symbols and their meanings are shown below ...
AND CLOTHING PROTECTION SHOULD BE WORN AT ALL TIMES. General Precautions Good laboratory practices should be followed and all procedures for use of the BD PrepMate™ Automated Accessory should be strictly observed. Failure to follow recommended procedures as outlined in this manual may compromise performance.
1.6.2 Mechanical Hazards The BD PrepMate™ Automated Accessory is a computer controlled device. There is a potential for injury and bodily harm from moving mechanical components when the instrument is in operation. A safety guard cover is provided to prevent accidental contact with any moving components.
TV reception or the functionality of other electrical devices. NOTE If it is necessary to interrupt operation of the instrument, press the STOP key on the user interface of the BD PrepMate™ Automated Accessory. 1.6.3 Electrical hazards The same precautions considered when using any electrical equipment should be observed with this instrument.
Operating sequence overview The following steps summarize the BD PrepMate™ Automated Accessory operating sequence. The user loads the BD PrepMate™ specimen rack with up to 12 specimens and processing components. The user loads the specimen rack into the BD PrepMate™ Automated Accessory and presses START.
The syringes are lowered so that their tips puncture the seals on the caps of the BD SurePath Liquid-based Pap Test preservative vials and stop near the bottom of each vial.To reduce the possibility of specimen contamination, the tops of vials are not removed during this process.
BD PrepMate™ Automated Accessory. HOME Use this key to return all BD PrepMate™ Automated Accessory components to their start-up positions. If a row is being processed when HOME is pressed the components do not return to the home position until that row is processed. This key is also used during power up to home the motors.
These keys do not have a role in the day-to-day use of the product. Do not use these keys. Operating the BD PrepMate™ Automated Accessory Please read this procedure carefully before operating. CAUTION Manually moving metal tray may cause instrument malfunction. Contact BD for assistance with moving the metal tray by hand.
1 Figure 5 – Loading the specimen rack Load the centrifuge tubes into the specimen rack. Insert each tube adjacent to its specimen vial, and then pipette 4 mL of BD Density Reagent into each tube. 4 mL density reagent...
Page 17
BD PrepStain™ slide tray is illustrated in Figure 7. The illustration identifies the location of Position 1 (row 1, column 1) in three places: the syringe, vial, and tube rows on the BD PrepMate™ specimen rack; the centrifuge rack rows; and the rows on the slide tray.
30 minutes to obtain optimum process results. CAUTION To maintain chain of custody, take care to properly orient the labeled, centrifuge tubes and specimen vials when loading the BD PrepMate™ specimen rack and when transferring the centrifuge tubes to centrifuge racks.
System Maintenance When a specimen rack processing cycle completes, an audible tone sounds. Replace the finished rack with the next rack to be processed, reset the number of rows if necessary, and press START. Inspect the centrifuge tubes to verify that the correct amount of sample has been transferred.
BD PrepMate™ Instrument User’s Manual Dispose of all wipes by placing them into a biohazard material container. As needed, wash the specimen racks in soap and warm water to remove any residue or buildup. This type of cleaning does not need to be performed after every run.
Operating Messages BD PREPMATE™ READY v x.x LOAD SPECIMEN RACK In this message, "x.x" indicates the current software version. When the normal start up sequence is complete, this message is displayed until a specimen rack is loaded into the instrument.
The mechanical draw action of the syringes is verified on each draw. This message is displayed if the draw action is insufficient. The last step of the BD PrepMate™ operating instructions is a visual verification of the draw volume. The absence of this error message indicates only that the mechanical draw action is correct.
The BD PrepMate™ Automated Accessory is designed to operate with no routine maintenance other than cleaning. Refer to System Maintenance for details. • Once every 12 months, preventive maintenance is required to be performed by BD authorized service personnel. Troubleshooting Instrument generated fault messages This section describes messages that indicate an operating fault condition.
Page 24
CLOSE VIAL HOLD-DOWN DOOR, PRESS RUN The vial hold-down door must be latched in the down position for the BD PrepMate™ Automated Accessory to run. Attached to this door are mechanical fingers that engage the top of the specimen vial so that the vial cannot be raised out of the rack when the syringe tip is removed from the vial.
Page 25
The last step of the BD PrepMate™ operating instructions is a visual verification of the draw volume. The absence of this error indicates only that the mechanical draw action is correct.
Page 26
Vial old-down door hinge screws loose NOTE If there is a loss of power to the BD PrepMate™ Automated Accessory, the processor does not retain memory. The operator will need to determine which step the process was in and restart testing...
Becton, Dickinson and Company Technical Support. 11.1 Warranty information The BD PrepMate™ Automated Accessory is warranted one year from the date of delivery. For repair or technical support information, contact your authorized distributor, or Becton, Dickinson and Company Technical Support.
12.5 Fusing The BD PrepMate™ has 2 fuses, located in the power entry module. All fuses are 5x20 mm and are 3.15 A @ 250 VAC time-delay, IEC 60127 rated. The product name plate is located on the rear panel with the power inlet and provides fuse information. Replacement fuses must be the...
Appendix: PrepMate™ Installation Instructions About these instructions These instructions describe the steps necessary to install the BD PrepMate™ Automated Accessory. The installer is expected to be either a BD PrepMate Automated Accessory user or an authorized representative of Becton, Dickinson and Company. It is also expected that the...
• Installing a power cord with incorrect polarity could create an electrical shock hazard or damage to the instrument. Hold down Specimen rack door latch Tray latch On/Off switch Figure A.1 - BD PrepMate™ Automated Accessory front view...
NOTE Use of a UPS (recommended but not supplied) between the power source and the BD PrepMate™ Automated Accessory is recommended to prevent interruption of the process in case of a power outage or brown out. The UPS should have a minimum rating of 150 VA for 5 minutes.
Page 33
Make sure that the plunger of each syringe is fully seated. NOTE The BD PrepMate™ Automated Accessory will not operate if a plunger is too high. NOTE For the purposes of this installation procedure, no live patient specimens are used.
The syringes are lifted and lowered back into the specimen rack. When a specimen rack processing cycle completes, an audible tone sounds. If the functions in steps 4 are not performed as described, shut down the BD PrepMate Automated Accessory and call Customer Service.
Need help?
Do you have a question about the PrepMate Automated Accessory and is the answer not in the manual?
Questions and answers