Download Print this page

Drucker Diagnostics HORIZON 6 Operation And Troubleshooting

Centrifuge
Hide thumbs Also See for HORIZON 6:

Advertisement

Quick Links

Centrifuge
Operation
and
Troubleshooting

Advertisement

loading
Need help?

Need help?

Do you have a question about the HORIZON 6 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Drucker Diagnostics HORIZON 6

  • Page 1 Centrifuge Operation Troubleshooting...
  • Page 2 Opposing tube holders must be equally loaded or empty or loaded with equally weighted samples. When loading only 3 tubes, they must be of equal weight. 6 TUBE CENTRIFUGES SETTINGS (1) Chemistry (2) Coag (PPP) (3) Urine 3,600 3,300 1,900 HORIZON 6 Time G-Force 1,800 xg 1,500 xg 500 xg druckerdiagnostics.com +1-866-265-1486 (U.S. only) support@druckerdiagnostics.com...
  • Page 3 TROUBLESHOOTING PROBLEM POSSIBLE CAUSE SOLUTION • Check line cord No power • Check circuit breaker on underside of centrifuge • Check wall outlet Rotate the lid knob fully clockwise before pressing the Lid knob is not completely closed ‘OPEN’ button The lid does not open Lid lock is active (Unlock timed out) Press the ‘OPEN’...
  • Page 4 Peel off the damaged sticker Wipe the surface with alcohol Replace the new front panel label, making sure that it is properly aligned in the channel on the molded panel For additional support, contact Drucker Diagnostics at 814-692-7661 Form 1048 Rev. A...