Denotes conformity to specific safety standards and regulations. FDA LISTED FDA Listed Denotes that the product has been properly listed with the FDA. Denotes conformity to quality standards and quality management ISO Certification systems. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 1...
General purpose laboratory centrifuge, intended for the density-based separation of fluids through centripetal acceleration. ARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
Running the centrifuge with an empty rotor will cause a motor error and will require a power cycle to fully reset (see Troubleshooting section on page 9). Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 3...
60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds. o Turn the lid knob counterclockwise and open the lid. The lid light will turn off. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
(Preset Lock). The Single Cycle Lock also prevents making changes to the parameters of the selected cycle. The Preset Lock allows selection of any saved cycle but prevents changing the parameters of saved cycles. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
Repeat. Save and Exit Press the GEAR button. The cycle displayed on the screen will be Programming Mode automatically saved. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 6...
Opposing tube holders must be equally loaded or empty or loaded with equally weighted samples. When loading only 3 tubes, they must be of equal weight. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 7...
Check the setting by following the instructions in the section on the desired value Changing Cycle Settings. If the preset is not the desired length, follow the procedure on the same page to change the preset time. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 9...
Page 12
Clicking noise during braking gets loud o Make sure that the screw in the center of the rotor is tight. Lid does not stay up o Tighten the center screw on the lid hinge. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 10...
In Inches: RCF or G-force = 0.00001118 x RCF or G-force = 0.0000284 x Rotor Radius (cm) x (RPM) Rotor Radius (in) x (RPM) Radius 4.5 in (11.5 cm) Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 11...
Page 14
Product Family: HORIZON Series (HORIZON 6 Flex FA) Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020 Protected by U.S. Patents #6,811,531, # 7,422,554, #D718,463, & #D734,489. Other Patents Pending RoHS FDA LISTED Compliant INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste.