Balancing Loads ..............................9 Care and Preventative Maintenance ........................9 Cleaning and Disinfection ..........................10 Troubleshooting ..............................11 General Specifications ............................13 Calculating the G-Force ............................. 14 Replacement Parts ............................14 Horizon 6 (DC) and Horizon 6 FA (DC) ........................14...
Denotes conformity to specific UK directives and regulations. FDA LISTED FDA Listed Denotes that the product has been properly listed with the FDA. Denotes conformity to quality standards and quality management ISO Certification systems. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 1...
Electrical Safety protection is provided by properly connecting the centrifuge to earth ground. Use only the manufacturer provided line cord and ensure that it is connected to a properly grounded power receptacle. Failure to do so will result in an electrical hazard. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 2...
Page 5
WARNING: Do not make modifications to or remove any hardware from rotor without prior authorization from Drucker Diagnostics. WARNING: Only use Drucker Diagnostics components in this centrifuge. Due to the lack of the possibility of human exposure, all Drucker centrifuges and accessories sold by Drucker Diagnostics, Inc.
General purpose laboratory centrifuge, intended for the density-based separation of fluids through centripetal acceleration. ARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
Turn on the power switch on the back of the centrifuge. BE SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY DISCONNECTION! Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 5...
To re-enable cycle selection, open lid, then press and hold the Unlock Cycle Selection UNLOCK button for 5 seconds. Three beeps will confirm that cycle selection is now unlocked. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 6...
Use a permanent marker to resist cleaning with a bleach solution. Erase with rubbing alcohol. For further protection, the label can be taped over with transparent tape after writing. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 8...
If cracks are discovered, replace immediately. Remove Accessories Before Moving: All tube holders, samples, and caps must be removed from the rotor chamber before transporting or storing the centrifuge to prevent damage and injury. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 9...
Check the setting by following the instructions in the section on the desired value Changing Cycle Settings. If the preset is not the desired length, follow the procedure on the same page to change the preset time. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 11...
Page 14
Clicking noise during braking gets loud o Make sure that the screw in the center of the rotor is tight. Lid does not stay up o Tighten the center screw on the lid hinge. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 12...
50% relative humidity 50% relative humidity at at 40 °C 40 °C Overvoltage Category Pollution Degree *The rotor and accessories are rated for the maximum rotation frequency shown. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 13...
6 (DC) 6 FA (DC) ORIZON ORIZON Part No. Description 7724037K Horizon 6 & Horizon 6 FA Foot, rubber (Pack of 4) 7786061 Horizon 6 Rotor, six-place, Horizontal 7786047 Horizon 6 FA Rotor, six-place, Fixed Angle 02-005-1-0010 Motor Assembly 02-006-0-0057...
Page 17
This operator’s manual is part number 03-0-0002-0308 Rev. A Product Family: Dash Series (Apex 6) Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020 Protected by U.S. Patents #6,811,531, # 7,422,554, #D718,463, #D734,489, & 10,994,285. Other Patents Pending FDA LISTED INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste.