7.8 Error Messages
If an error appears, it will be displayed in the GUI as a pop-up message that contains the error
description and the time stamp, for example:
Additionally, in the status bar a notification is displayed:
Click Close to remove the error message from the GUI. Click Clear All to reset the MTD415x
internal error register.
Note
Closing an error message does not clear the MTD415x error register, it just makes the
message disappear from the GUI foreground. In this case, the status bar notification remains.
Only Clear All will reset the error register and delete the notification. Clearing the error register
is mandatory to resume normal operation. Alternatively, switch SW1 off and on again.
Example
The TEC connection was interrupted. The TEC current is switched off, and the appropriate
error message is displayed.
Reconnect the TEC. The MTD415x will resume operation only after clicking to Clear All or by
switching OFF and ON again the
© 2016 Thorlabs GmbH
enable switch
(8).