(refer to as instructed on Rocker’s cleaning solution). Note: Inflammable or low flash point detergents should be never used. As the energy generated by the cavitation phenomenon will be converted to heat and kinetic energy that would increase the solution temperature and make low flash-point solution flammable.
Unpacking Carefully check if the shipping carton is in good condition before unpacking. Please check all supplied items according to standard package list. If you have any questions, please keep serial no. & shipping carton and contact your local dealer for help. The complete package contains: Ultrasonic Cleaner x 1 Power Cord x 1...
Operation A. Installation Please place the unit on a ventilated and clean table surface. Check voltage specified on the nameplate at the back of the unit. Ensure if it matches the line voltage on your location. Plug in the power. Pour the formulated cleaning solution into the chamber up to the operating level after the object being cleaned is placed in.
Page 5
Change the last settings: 1. Cleaning Mode (Degas indicator unlighted) 1.1 Change the cleaning time: press (2) then (10) starts flashing => press (3) or (4) to increase or decrease the setting value => press (6) => start cleaning. 1.2 Change into Degas Mode: press (2) for 3 times =>...
Troubleshooting Conditions Possible Causes Solutions Unconnected power plug Plug in power exactly Failure to start-up Invalid power outlet Replace the power outlet Fuse burned out Contact service technician Too much bubble coming from cleaning solution ...