VLP-100B Series Label Printing Scale Operation Manual REV:18 13. CALIBRATION (Dealer only) It allows selecting Normal calibration、Linear calibration、Zero calibration to calibrate the scale. 13.1 Normal Calibration in lb Mode ▪ Press key to enter setup menu. ▪ Using numeric keys 5、2、3 to select calibration unit. It may be asked to input the password,default is1234 5.Service>...
Page 3
VLP-100B Series Label Printing Scale Operation Manual REV:18 4 Press key to confirm. 5 Place the test weight on the platter. 6 Press key to confirm. The display will be show “Please empty the scale”. 7 Remove the test weight from the scale.
Page 4
VLP-100B Series Label Printing Scale Operation Manual REV:18 13.2 Normal Calibration in kg Mode 5.service> 2.calibration parameters > 5.Calibration> 1.Normal Calibration ▪ Before access to the parameter,it need to press calibration switch,locating back side of indicator in approval version.It may be asked to input the password,default is1234 ▪...
Page 5
VLP-100B Series Label Printing Scale Operation Manual REV:18 4 Press key to confirm. 5 Place the test weight on the platter. 6 Press key to confirm. The display will be show “Please empty the scale”. 7 Remove the test weight from the scale.