DN:KD-5811BT-SMSY01 V1.0 Operation Guide IMPORTANT INFORMATION NORMAL BLOOD PRESSURE FLUCTUATION All physical activity, excitement, stress, eating, drinking, smoking, body posture and many other activities or factors (including taking a blood pressure measurement) will influence blood pressure value. Because of this, it is mostly unusual to obtain identical multiple blood pressure readings.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide INTENDED USE Fully Automatic Electronic Sphygmomanometer is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide SPECIFICATIONS 1. Product name: Wireless Blood Pressure Monitor 2. Model: KD-5811BT 3. Classification: Internally powered, Class II, Type BF applied part, IP21, No AP or APG, Continuous operation Machine size: Approx.139.4mm×93.8mm×43.4mm( 5 1/2″x 3 11/16″x 1 23/32″) Cuff circumference: 22cm-42cm(8 21/32 ″...
Page 6
DN:KD-5811BT-SMSY01 V1.0 Operation Guide Measuring on same arm for each measurement. Please always relax a minimum moment of 1 to 1.5 minutes between measurements to allow the blood circulation in your arm to recover. Prolonged over-inflation (cuff pressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause ecchymoma of your arm.
Page 7
DN:KD-5811BT-SMSY01 V1.0 Operation Guide monitor, such as input: 100-240V, 50/60Hz, 0.2A; output: DC 5V, 1A. Please note that the monitor jack size: hole Φ5.5mm, center pin Φ2.0mm. 17. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide SETUP AND OPERATING PROCEDURES 1. BATTERY LOADING AND AC ADAPTER LOADING BATTERY LOADING a. Open battery cover at the back of the monitor. b. Load four “AA” size batteries. Make sure the batteries are inserted according to the positive and negative marks("+"and"-") printed in the battery housing.
Page 9
DN:KD-5811BT-SMSY01 V1.0 Operation Guide c. In Clock and Date Adjustment Mode , the time format will blink at first , see picture2-1 .The default time format is 24h and the default clock and date is 2023-1-1 1:00. d. Press the button “START/STOP” repeatedly, the year (first usage: default is 2023, range is 2023~2099), month, day, hour and minute will blink in turn, see picture 2- 2&...
DN:KD-5811BT-SMSY01 V1.0 Operation Guide 2.2 Table 1 instructs the conversion relations between 24 hour format and 12 hour format. Table 1 24 hour format 12 hour format 24 hour format 12 hour format 0:00 12:00 AM 12:00 12:00 PM 1:00...
DN:KD-5811BT-SMSY01 V1.0 Operation Guide Picture 4-1 Picture 4-2 Picture 4-3 Note: Please refer to the cuff circumference range in “SPECIFICATIONS” to make sure that the appropriate cuff is used. Measure on the same arm each time. Do not move your arm, body, or the monitor and do not move the rubber tube ...
DN:KD-5811BT-SMSY01 V1.0 Operation Guide b. Then the current memory bank ( , ) is displayed. If the monitor has stored results, the LCD will momentarily display the most recent one. If no result has been stored, zero will appear on LCD. Press “MEM1” or “MEM2”button to change over to other bank.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide Picture 7-1 Picture 7-2 Picture 7-3 Picture 7-4 Picture 7-5 Picture 7-6 Picture 7-7 Picture 7-8 d.Press current bank MEM button again, the most recent result will be displayed with date and time stamp. See picture 7-6. Irregular heartbeat symbol (if any) and blood pressure classification indicator will blink at the same time.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide with your physician for accurate assessment, and never change your treatment by yourself. Note: It is not intended to provide a basis of any type of rush toward emergency conditions/diagnosis based on the number scheme and that the number scheme is meant only to discriminate between the different levels of blood pressure.
DN:KD-5811BT-SMSY01 V1.0 Operation Guide It is inappropriate for people with serious arrhythmia to Irregular heartbeat (arrhythmia) use this Electronic Sphygmomanometer. 12. TROUBLESHOOTING (2) PROBLEM POSSIBLE CAUSE SOLUTION LCD shows Low Battery Change the batteries low battery symbol LCD shows Pressure system is unstable before “Er 0”...
DN:KD-5811BT-SMSY01 V1.0 Operation Guide MAINTENANCE Do not drop this monitor or subject it to strong impact. Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the monitor. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, In c and any use of such marks by ANDON HEALTH CO., LTD. is under license. Other trademarks and trade names are those of their respective owners.
Equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the KD-5811BT, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.”...
Page 19
DN:KD-5811BT-SMSY01 V1.0 Operation Guide Discharge ±2kV, ±4kV, ±8kV, ±15kV air Radiated RF EM IEC 61000-4-3 10V/m field 80MHz-2.7GHz 80% AM at 1kHz Proximity fields IEC 61000-4-3 Refer to table 3 from RF wireless communications equipment Rated power IEC 61000-4-8 30A/m...
Page 20
DN:KD-5811BT-SMSY01 V1.0 Operation Guide fields 80%AM at 1kHz Voltage dips IEC 61000-4-11 0% U ; 0.5 cycle At 0º, 45º, 90º, 135º, 180º, 225º, 270 º and 315º 0% U ; 1 cycle 70% U ; 25/30 cycles Single phase: at 0º...