Minitron - Operating manual
Options
4.7.2
Using the Cable Pass-through
Page 46 of 128
WARNING
If an external device that is supplied with mains voltage is operated
on the shaking table whilst it is operating, there is a risk that the
cable might break or get jammed. This poses a risk of death by
electrocution.
−
Any mains voltage operated devices may only be used when
the shake function is switched off (e.g. in conjunction with the
"Temperature" parameter.)
−
Any mains voltage operated devices may only be used in con-
junction with the "Humidity" parameter if the device is ade-
quately protected against humidity. Observe the permissible
environmental conditions for operating the device.
NOTICE
To avoid damages when installing the cables and/or hoses in the
incubation chamber, make sure that they do not rub and do not get
jammed.
To correct insert cables and hoses into the incubation chamber, pro-
ceed as follows:
1.
Loosen the knurled screws on the cable pass-through frame.
2.
Remove the cable pass-through.
3.
Push the blue interior part out of the frame.