Indications for Use/Intended Use Warning The LCD Monitor is intended to provide 4K 2D color This unit has no power switch. video displays of surgical and other medical To disconnect the main power, unplug the power images. plug. When installing the unit, incorporate a readily Regulatory Information relevant under accessible disconnect device in the fixed wiring, or the European Medical Device Regulation...
Trade Name : Sony Corporation Model : LMD-XH550N Responsible party : Sony Electronics Inc. Storage and transport temperature Address : 16535 Via Esprillo, This symbol indicates the acceptable San Diego, CA 92127 U.S.A.
Page 4
Important safeguards and notices for use in the medical environments 1. All devices connected to the unit must be certified or compliant according to IEC 60601-1, IEC 60950-1, IEC 62368-1 and IEC 60065 standards and other IEC/ISO standards applicable to the devices. 2.
Page 5
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550N. Otherwise, degradation of the performance of this equipment could result. • If the LMD-XH550N will be used adjacent to or stacked with other equipment, normal operation of the LMD-XH550N under such configurations should be verified via observation.
Page 6
Guidance and manufacturer’s declaration – electromagnetic immunity The LMD-XH550N is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-XH550N should assure that it is used in such an environment. IEC 60601 Immunity test Compliance level Electromagnetic environment –...
Page 7
Guidance and manufacturer’s declaration – electromagnetic immunity The LMD-XH550N is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-XH550N should assure that it is used in such an environment. IEC 60601 Compliance Immunity test Electromagnetic environment –...
Page 8
If the measured field strength in the location in which the LMD-XH550N is used exceeds the applicable RF compliance level above, the LMD-XH550N should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the LMD-XH550N.
Page 9
The LMD-XH550N is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-XH550N. Otherwise, degradation of the performance of this equipment could result.
Page 10
Caution Warning When you dispose of the unit or accessories, you To prevent injury, if mounting the unit using a must obey the laws in the relative area or country mounting arm, wall fixture, or other mounting and the regulations in the relative hospital device prepared by the customer, mount the unit regarding environmental pollution.
Page 11
For the customers in the U.S.A. SONY LIMITED WARRANTY - Please visit www.sony.com/psa/warranty for important information and complete terms and conditions of Sony’s limited warranty applicable to this product. For the customers in Canada SONY LIMITED WARRANTY - Please visit www.sony.com/psa/warranty for important...
Preset menu .........34 Table of Contents Troubleshooting ..........34 Error Messages ..........35 Specifications ..........35 Precaution ............13 Dimensions ............41 On Safety ...........13 Licenses ............42 On Installation ...........13 Safety precautions for using this unit ..13 Precautions when connecting this unit to medical equipment .........13 To prolong the life of the unit ....13 Cautions for RESPONSIBLE ORGANIZATION...
When connecting the unit to medical Precaution equipment, refer to “Precautions when connecting this unit to medical equipment”. On Safety Precautions when connecting this unit to medical equipment Operate the unit only with a power source as specified in the “Specifications” section. Before using this unit for medical purposes, be ...
update of equipment connected to the IT- addition, over a long period of use, because of NETWORK; and the physical characteristics of the liquid crystal upgrade of equipment connected to the IT- display, such “stuck” pixels may appear NETWORK.
On Burn-in On Security For LCD panel, permanent burn-in may occur if SONY WILL NOT BE LIABLE FOR DAMAGES OF still images are displayed in the same position on ANY KIND RESULTING FROM A FAILURE TO the screen continuously, or repeatedly over IMPLEMENT PROPER SECURITY MEASURES ON extended periods.
If you have any questions about this unit, contact your authorized dealer. 1) A.I.M.E. is a registered trademark of Sony Corporation. Control panel Assigns functions frequently used during an operation to buttons on the front surface of the monitor.
Page 17
Flat surface for better maintenance The design allows the user to easily wipe liquids and gel off the LCD panel and control buttons – ensuring a high standard of disinfection and cleanliness.
Location and Function of Parts and Controls Front Panel Power indicator When you press PORT A or PORT B when it is lit in white, an image from the input Indicator Operating State connector assigned to the PORT A/B is displayed.
CUSTOM buttons Turns on or off the assigned function. You can adjust the assigned function by pressing the / buttons. (Refer to the Custom Button of the System Configuration menu on page 33 and of the Initial Setup menu on page 34.) The following functions are assigned in the default setting.
Rear Panel To use the connectors on the rear panel, remove the cable cover. For details on the cable cover, see page 24. Cable cover (L) cover to the right. The cable of the 12G-SDI 1 When connecting or disconnecting the cable of input/output connector can be connected or the connector on the rear panel, remove this disconnected without removing the cable cover...
HDMI cable holder the communication speed or quality of 10BASE- Secures the HDMI cable (Ø7 mm or less). T/100BASE-TX. HDMI input connector Close Inputs the HDMI signals. Use the Premium High Speed cable that is shorter than 3 m (meters) with the cable-type Cable logo.
Be sure to connect equipment or cables 12V 2.5A (DC output) connector specified by Instructions for Use of this monitor Outputs 12 volts for connected external to the 12G-SDI/3G-SDI output connectors. If equipment. you connect unspecified equipment or cables, ...
Preparation Connecting AC input connector Make sure that the (On)/ (Standby) switch is set to (Standby). AC power cord Remove the cable cover (L). Slide down the cable cover (L) while sliding Plug holder for the the slide locks (2 pcs), which are located at supplied AC power cord the bottom of the monitor, in the direction of the arrow.
Attach the cable cover (L). Turning on the Monitor/Switching Input Slide the cable cover (L), while sliding the Settings slide locks (2 pcs), which are located at the bottom of the monitor, in the reverse Connect the plug of the AC power cord to an direction of step2.
Area Color Temperature Initial Setting North America Latin America - Argentina Latin America - Paraguay Latin America - Uruguay When you turn on the unit for the first time after Latin America - Other purchasing it, select the area and language Africa, Australasia where you intend to use this unit from among Europe, Middle-East...
Using the Menu The unit is equipped with an on-screen menu for making various adjustments and settings such as picture control, input setting, setting change, etc. Press the / button to select the tab. The selected tab is shown in blue, and setting items of the selected tab appear.
About the control navigation Adjustment Using the Menus Depending on the state, the operation buttons of the unit light as shown below: White light: Operable state. Green light: Operating state. Items Off: Unable to operate. The screen menu of this monitor consists of the following items.
Input Name Menu Setting AUX IN Setting Chroma Adjusts color intensity. The higher the Power Supply setting, the greater the intensity. The lower the setting, the lower the intensity. System Configuration (page 31) Brightness Adjusts brightness. Control Lock Contrast Adjusts contrast. OSD Setting A.I.M.E.
Menu Setting Menu Setting RGB Range Select the RGB signal range from 1) Signals in resolutions of 3840 × “Auto,” “Limited,” or “Full.” If you set to 2160 or 4096 × 2160 are listed “Auto,” this item is set to “Limited” below: when inputting video signals, and 4:4:4 RGB/YCbCr-50P/60P-8bit...
Menu Setting Menu Setting Sets the sub screen position for the 2 Input Name Set the name of the 12G-SDI 1, 12G-SDI or 3 screens display (POP). 2, 3G-SDI, Display Port, HDMI, DVI-D input connector. Pattern Skip Sets the pattern that is skipped when ...
Page 32
Menu Setting Menu Setting Power On Setting Input Connector Name Power On Mode Select the setting when the monitor is turned on from the following settings. Select Port Control Lock Last: The setting when the monitor was last turned off. ...
Menu Setting Menu Setting Remote Backlight Control Off: Turns off backlight control. Remote Mode Sets the remote function when the Standard: REMOTE connector is connected to the Sets the backlight to standard optional Foot Switch FS-24. brightness mode. ...
Menu Setting Troubleshooting PORT A HD Select the clipping size of the HD Clipping Size/ signal from “Normal,” “4:3,” “5:4,” or “V PORT B HD Full” when displaying 2 or 3 screen Clipping Size displays. This section may help you isolate the cause of a Custom Button Assigns the function to the CUSTOM 1 problem and as a result, eliminate the need to...
Error Messages Specifications Picture performance When the following messages appear on the screen, turn off the power and contact an LCD panel TFT Active Matrix authorized dealer. Pixel efficiency 99.99% Messages Description Viewing angle (panel specification) Fan Error A malfunction occurs in the fan. 89°/89°/89°/89°...
Page 36
700 hPa to 1 060 hPa WHATSOEVER. Accessories supplied SONY WILL NOT BE LIABLE FOR CLAIMS OF AC power cord (1) ANY KIND MADE BY USERS OF THIS UNIT OR Plug holder for the supplied AC MADE BY THIRD PARTIES.
Pin assignment 12V 2.5A (DC output) connector SERIAL REMOTE (RS-232C) connector D-sub 9-pin, female Pin number Signal Pin number Signal 12 V REMOTE (stereo mini jack) connector Pin number Signal On (short to sleeve) Off (open) 5V 2.0A (DC output) connector Pin number Signal...
Page 38
Available signal formats The unit is compatible with the signal systems shown below: Signal format 12G-SDI 1 12G-SDI 2 3G-SDI SDI Single Link SD-SDI 720×487/59.94i 4 : 2 : 2 YCbCr 10bit × × 720×576/50i 4 : 2 : 2 YCbCr 10bit ×...
When mounting the unit to objects such as Dimensions movable mounting arms, if excessive force is applied, the mounting part may become damaged, and in the worst case, it may cause injury or damage to the unit due to the unit Front falling.