SETTING UP THE INSTRUMENT

1. Connect the plug on the AC adapter to the port on the back of the ColorFlex L2.
2. Plug the power cord on the AC Adapter into a wall outlet.
3. Turn on the ColorFlex L2 using the switch on the back of the instrument.
4. The Essentials software will launch automatically and prompt First Time Setup dialog to setup
Languages, Region, Date and Time. Click DONE when finished.
5. After the setup above, a Welcome Wizard will be prompted to introduce its features. Please
follow the prompts to familiarize yourself with the software. Once completed, your instrument
will be ready for standardization. You can always click the HunterLab icon in the top right corner
to run the Wizard again.
6. (Optional) Connect a monitor to the ColorFlex L2 using an HDMI cable.
B
STANDARDIZATION
1. Press the STANDARDIZATION button on the screen.
2. Open the Standards Case and place the Reflectance Black Glass at the port. Press
the MEASURE button.
3. When prompted, remove the Reflectance Black Glass and place the Instrument Standard (White
Tile) on the port. Press MEASURE.
AC ADAPTER PORT
SWITCH