QA3™ v2.0 OPHTHALMIC DESIGN PATIENT STRETCHER SYSTEM
MPN
Name
21118
QA3™ Ophthalmic Stretcher
Serial number 28661 onwards only - manufactured from 2024.
Frequency of service
A QA3™ Ophthalmic Stretcher variant is to be serviced once annually.
Lifetime
The life expectancy of a QA3™ Ophthalmic Stretcher variant is 10 years from date of introduction to
clinical use, dependent on the level of care and maintenance. The performance of this device may reduce
once the life expectancy has been reached and exceeded.
No component needs compulsory replacement during the lifetime of the device.
Day-to-Day maintenance
Before use, ensure all stretcher functions operate to their full range of movement and that all
components disengage, re-engage and lock correctly. Also visually inspect the stretcher for any loose or
damaged parts, foreign bodies caught in the castors and hydraulic fluid leakage.
If the stretcher is damaged or faulty it must be taken out of use with immediate effect and the fault
reported to Anetic Aid or maintenance department. The stretcher must not be used until the damage or
fault has been repaired.
Regulations
The following regulations will be adhered to as part of the servicing activities:
Health and Safety at Work Regulations 1999 (management regulations)
The Health and Social Care Act 2008 (Regulated Activities) Regulations 2014
TECHNICAL DOCUMENTATION
SERVICING INFORMATION
TECH 01.0 – 800.16
Page 1 of 4

Advertisement

loading
Need help?

Need help?

Do you have a question about the QA3 v2.0 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Anetic Aid QA3 v2.0

  • Page 1 If the stretcher is damaged or faulty it must be taken out of use with immediate effect and the fault reported to Anetic Aid or maintenance department. The stretcher must not be used until the damage or fault has been repaired.
  • Page 2 TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 01.0 – 800.16 Guidance documents The following guidance documents will be referenced as part of the servicing activities: • 992015 – QA3 Powered Function Stretchers Instruction for Use. • Medicines & Healthcare products Regulatory Agency: Managing Medical Devices Guidance for health and social care organisations.
  • Page 3 TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 01.0 – 800.16 SERVICE SCHEDULE 1. Initial set-up & inspections 1.1. Ensure all four castors are trailing away from the foot-end. 1.2. Apply brakes. 1.3. Press the On / Off button to turn the device on. 1.4.
  • Page 4 TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 01.0 – 800.16 7. Tilt actuator, Column & Internal Inspection 7.1. Check fixings & condition of tilt actuator. 7.2. Check condition of column stability. 7.3. Visual inspection of internal fixings. 7.4. Visual inspection of internal welding. 8.

This manual is also suitable for:

21118