Anetic Aid QA4 v2.0 Service Manual

Design mobile surgery system

Advertisement

Quick Links

TECHNICAL DOCUMENTATION
QA4™ v2.0 DESIGN MOBILE SURGERY SYSTEM
MPN
Name
21300
QA4™ Mobile Surgery System
Serial number 551 to 3450 only - manufactured between 2010 and 2020.
GMDN
33152 Universal operating table, electromechanical
Frequency of service
A QA4™ Mobile Surgery System is to be serviced once annually.
Lifetime
The life expectancy of a QA4™ Mobile Surgery System is 10 years from date of introduction to
clinical use, dependent on the level of care and maintenance. The performance of this device may
reduce once the life expectancy has been reached and exceeded.
The battery requires replacement during the lifetime of the device – once every three years.
Day-to-Day maintenance
Before use, ensure all stretcher functions operate to their full range of movement and that all
components disengage, re-engage and lock correctly. Also visually inspect the stretcher for any
loose or damaged parts, foreign bodies caught in the castors and hydraulic fluid leakage.
If the stretcher is damaged or faulty it must be taken out of use with immediate effect and the
fault reported to Anetic Aid or maintenance department. The stretcher must not be used until the
damage or fault has been repaired.
Regulations
The following regulations will be adhered to as part of the servicing activities:
Health and Safety at Work Regulations 1999 (management regulations)
The Health and Social Care Act 2008 (Regulated Activities) Regulations 2014
SERVICING INFORMATION
TECH 02.0 – 800.2
Page 1 of 5

Advertisement

loading
Need help?

Need help?

Do you have a question about the QA4 v2.0 and is the answer not in the manual?

Questions and answers

Summary of Contents for Anetic Aid QA4 v2.0

  • Page 1 If the stretcher is damaged or faulty it must be taken out of use with immediate effect and the fault reported to Anetic Aid or maintenance department. The stretcher must not be used until the damage or fault has been repaired.
  • Page 2: Guidance Documents

    TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 02.0 – 800.2 Guidance documents The following guidance documents will be referenced as part of the servicing activities: • 992010 QA4 Powered Mobile Surgery System Instructions for Use. • Medicines & Healthcare products Regulatory Agency: Managing Medical Devices Guidance for health and social care organisations.
  • Page 3 TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 02.0 – 800.2 SERVICE SCHEDULE 1. Initial Set-up & Inspections 1.1. Ensure all four castors are trailing away from the foot-end. 1.2. Apply brakes. 1.3. Lift all cotsides into raised position. 1.4. Detach the mains lead (if connected). 2.
  • Page 4: Lateral Tilt

    TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 02.0 – 800.2 8. Leg Section 8.1. Check fixings & condition of leg section board. 8.2. Remove leg section and check condition of locating pins. 8.3. Check fixings & condition of locating blocks & release buttons. 8.4.
  • Page 5 TECHNICAL DOCUMENTATION SERVICING INFORMATION TECH 02.0 – 800.2 End of Document Date of Change Issue Brief Description of Change Signature Replaces document May 2024 reference 95703D Page 5 of 5...

This manual is also suitable for:

21300