Emerson Flexim FLUXUS G722ST-HT Operating	 Instruction

Emerson Flexim FLUXUS G722ST-HT Operating Instruction

Ultrasonic flowmeter
Table of Contents

Advertisement

Quick Links

Operating Instruction
UMFLUXUS_G722ST-HTV2-2-0-0EN
AN 26754
2024-03-01
Flexim FLUXUS G722ST-HT
Ultrasonic Flowmeter

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Flexim FLUXUS G722ST-HT and is the answer not in the manual?

Questions and answers

Summary of Contents for Emerson Flexim FLUXUS G722ST-HT

  • Page 1 Operating Instruction UMFLUXUS_G722ST-HTV2-2-0-0EN AN 26754 2024-03-01 Flexim FLUXUS G722ST-HT Ultrasonic Flowmeter...
  • Page 3: Table Of Contents

    FLUXUS G722ST-HT Table of contents Table of contents Introduction ................7 Safety instructions .
  • Page 4 Table of contents FLUXUS G722ST-HT Measurement ................81 Parameter input .
  • Page 5 FLUXUS G722ST-HT Table of contents Settings ................144 19.1 Dialogs and menus .
  • Page 6 Table of contents FLUXUS G722ST-HT 2024-03-01, UMFLUXUS_G722ST-HTV2-2EN...
  • Page 7: Introduction

    1 Introduction FLUXUS G722ST-HT Introduction This operating instruction has been written for users operating the ultrasonic flowmeter FLUXUS. It contains important information about the measuring equipment, how to handle it correctly, and how to avoid damages. Read the safety instructions carefully. Make sure you have read and understood this operating instruction before using the measuring equipment.
  • Page 8: Safety Instructions

    2 Safety instructions 2.1 General safety instructions FLUXUS G722ST-HT Safety instructions General safety instructions Prior to any work, read the operating instruction carefully and in full. Failure to comply with the instructions, in particular with the safety instructions, poses a risk to health and can lead to material damages.
  • Page 9: Intended Use

    2 Safety instructions FLUXUS G722ST-HT 2.2 Intended use Intended use The measuring equipment is intended for the measurement of fluid properties in closed pipes. The connected transducers measure the average flow velocity of the fluid. The transmitter uses these values to calculate the sought quantities, e.g., volumetric flow rate and mass flow rate. Through comparison with the values stored in the transmitter further physical quantities can be determined.
  • Page 10: Safety Instructions For Transport

    2 Safety instructions 2.7 Safety instructions for transport FLUXUS G722ST-HT marked as a disconnecting device for the transmitter. It should be located near the transmitter. If the transmitter is used in an explosive atmosphere, the equipment switch has to be installed outside the explosive atmosphere. If this is not possible, it has to be installed in the least hazardous area.
  • Page 11: General Principles

    3 General principles FLUXUS G722ST-HT 3.1 Measurement principle General principles In the ultrasonic flow measurement, the flow velocity of the fluid in a pipe is determined. Further physical quantities are derived from the flow velocity and from additional physical quantities, if necessary. Measurement principle 3.1.1 Terms...
  • Page 12 3 General principles 3.1 Measurement principle FLUXUS G722ST-HT 3.1.2 Measurement of the flow velocity The flow velocity of the fluid is measured using the correlation principle. 2 pairs of ultrasonic transducers of the same type are mounted one after the other at a distance Δx on the pipe which is completely filled with the fluid. The transducer pairs form the measuring barriers A and B.
  • Page 13: Measurement Arrangements

    3 General principles FLUXUS G722ST-HT 3.2 Measurement arrangements Measurement arrangements 3.2.1 Terms The transducers of a measuring barrier are mounted on opposite sides of the pipe. Sound path The distance covered by the ultrasonic signal of a transducer pair after crossing the pipe once. Distance between the measuring barriers The measuring barrier distance is measured between the inner edges of the emitters and/or receivers of measuring barrier A and B.
  • Page 14: Acoustic Penetration

    3 General principles 3.3 Acoustic penetration FLUXUS G722ST-HT Sound beam plane Plane containing the sound paths of the measuring barriers. Fig. 3.4: 2 sound paths in 1 plane Acoustic penetration The pipe has to be acoustically penetrable at the measuring point. The acoustic penetration is given when pipe and fluid do not attenuate the sound signal so strongly that it is completely absorbed before reaching the second transducer.
  • Page 15 3 General principles FLUXUS G722ST-HT 3.4 Undisturbed flow profile Tab. 3.1: Recommended distance from disturbances d – inner pipe diameter at the measuring point l – recommended distance between disturbance and transducer position disturbance: 90° elbow inlet: l ≥ 10 d outlet: l ≥...
  • Page 16 3 General principles 3.4 Undisturbed flow profile FLUXUS G722ST-HT Tab. 3.1: Recommended distance from disturbances d – inner pipe diameter at the measuring point l – recommended distance between disturbance and transducer position disturbance: 45° double elbow inlet: l ≥ 15 d outlet: l ≥...
  • Page 17: Product Description

    4 Product description FLUXUS G722ST-HT 4.1 Measuring system Product description Measuring system The measurement system consists of the transmitter, 2 pairs of ultrasonic transducers and the pipe on which the measurement is carried out. Fig. 4.1: Example of a measurement arrangement 1 –...
  • Page 18 4 Product description 4.2 Handling concept FLUXUS G722ST-HT Tab. 4.1: Description of the menus menu description Parameters input of sensor, pipe and fluid parameters Installation input of measuring point specific parameters measurement start Start measurement measured value display Show measurement Stop measurement measurement stop Inputs...
  • Page 19: Display

    4 Product description FLUXUS G722ST-HT 4.3 Display Display Structure Fig. 4.3: Menu Parameters (example) 1 – menu 2 – menu item currently edited 3 – area for scroll lists, selection fields or input fields Tab. 4.2: Navigation horizontal scroll list vertical scroll list input fields •...
  • Page 20: Keyboard

    4 Product description 4.4 Keyboard FLUXUS G722ST-HT Keyboard The keyboard has 15 keys, including 3 function keys: ENTER, and C. Some keys have multiple functions. They can be used to enter data, to navigate through scroll lists as well as to execute special functions (e.g., reset of totalizers).
  • Page 21: Transport And Storage

    5 Transport and storage FLUXUS G722ST-HT 5.1 Transport Transport and storage Caution! When packaging, the transmitter can fall down. There is a danger of crushing body parts or damaging the measuring equipment. → Secure the transmitter against falling during packaging. →...
  • Page 22: Mounting

    6 Mounting FLUXUS G722ST-HT Mounting Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Installation, connection and start-up by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 23: Transmitter

    6 Mounting FLUXUS G722ST-HT 6.1 Transmitter Transmitter 6.1.1 Opening and closing the housing 6.1.1.1 Opening Caution! Possible danger by opening the equipment at improper ambient conditions Open the transmitter in safe ambient conditions only (e.g., air humidity < 90 %, no conductive pollution, no explosive atmosphere).
  • Page 24: Installation Of The Transmitter

    6 Mounting 6.1 Transmitter FLUXUS G722ST-HT Transmitter with aluminum housing • Close the housing cover pressing it slightly upwards on the side of the screws. • Hand-tight all screws of the transmitter housing. Fig. 6.2: Transmitter 1 – screws 6.1.2 Installation of the transmitter Notice! Install the equipment in a shady place, the cable glands facing downward.
  • Page 25 6 Mounting FLUXUS G722ST-HT 6.1 Transmitter Transmitter with aluminum housing • Fix the wall mount (2) with 3 countersunk-head screws (3) firmly to the wall. • Attach the transmitter (1) to the lower hooks (7) of the wall mount. • Fix the transmitter to the wall mount by locking the stop bolt (4) into the destined hole and by firmly tightening the screw (5). Fig.
  • Page 26: Pipe Mounting

    6 Mounting 6.1 Transmitter FLUXUS G722ST-HT 6.1.2.2 Pipe mounting Notice! The pipe has to be sufficiently stable to withstand the pressure exerted by the transmitter and the shackles. Transmitter with stainless steel housing Installation on a 2'' pipe • Position the transmitter and the shackle (1) on the pipe. •...
  • Page 27 6 Mounting FLUXUS G722ST-HT 6.1 Transmitter Transmitter with aluminum housing Notice! For the mounting on pipes the pipe mounting kit, consisting of 2 band clamps, 2 countersunk-head screws and 2 washers, is needed. The distance of the band clamps has to be identical to the distance of the holes on the wall mount. •...
  • Page 28: Transducers

    6 Mounting 6.2 Transducers FLUXUS G722ST-HT Transducers Mounting Caution! Warning of severe injuries from hot or cold components Touching hot or cold components can lead to severe injuries (e.g., thermal damage). → Any mounting, installation or connection work has to be concluded. →...
  • Page 29: Pipe Preparation

    6 Mounting FLUXUS G722ST-HT 6.2 Transducers 6.2.1.2 Pipe preparation Caution! Contact with grinding dust This may result in injuries (e.g., breathing difficulties, skin reactions, eye irritations). → Wear the required personal protective equipment. → Observe the applicable rules. Important! The pipe has to be sufficiently stable to withstand the pressure exerted by the transducers and the fasteners. Notice! Observe the selection criteria of pipe and measuring point.
  • Page 30: Installation Of The Transducers

    6 Mounting 6.2 Transducers FLUXUS G722ST-HT 6.2.2 Installation of the transducers 6.2.2.1 Orientation of the transducers and determination of the distance between the transducers and measuring barriers Observe the orientation of the transducers. If the transducers of a transducer pair have been mounted properly, the engravings on them form an arrow pointing in the flow direction.
  • Page 31 6 Mounting FLUXUS G722ST-HT 6.2 Transducers 6.2.2.2 Transducer arrangement The transducers can be arranged in the mounting rails in different ways: Fig. 6.9: Transducer arrangement in mounting rails all transducers in one pair of rails a small, Δx small each transducer pair in its own pair of rails: a small, Δx large transducers distributed over 3 rail positions: a large, Δx large...
  • Page 32 6 Mounting 6.2 Transducers FLUXUS G722ST-HT 6.2.2.4 Mounting with Variofix L (PermaRail) Mounting For measurements in diagonal arrangement, the transducer mounting fixtures are mounted on opposite sides of the pipe. Mounting the transducer pairs in one pair of rails is not possible when using Variofix L. In the following, the installation of 2 pairs of rails in diagonal arrangement is described (each transducer pair in its own pair of rails).
  • Page 33 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Step 2: Fixation of the clasps to the tension straps • Select the installation instruction of the supplied clasp: Band clamp clasp The clasp is fixed to the tension strap, see Fig. 6.12. Quick release clasp The clasp is fixed to the tension strap, see Fig.
  • Page 34 6 Mounting 6.2 Transducers FLUXUS G722ST-HT Step 3: Fixation of the tension strap to the pipe One tension strap is fixed to the pipe using 2 opposite tension strap clamps. The second tension strap will be mounted later. Fig. 6.15: Tension strap with clamp and metal spring on the pipe 1 –...
  • Page 35 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Fig. 6.18: Tension strap with band clamp clasp 1 – clasp screw • Place the tension strap around the pipe and insert it into the clasp, see Fig. 6.17. • Tighten the tension strap. •...
  • Page 36 6 Mounting 6.2 Transducers FLUXUS G722ST-HT Fig. 6.20: Ratchet clasp with tension strap 1 – sense of rotation 2 – edge 3 – lever 4 – clasp screw with slot Step 4: Fixation of the rail to the pipe • Place the tension strap clamp (2) in the rail, see Fig. 6.21. Observe the orientation of the tension strap clamp. •...
  • Page 37 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Band clamp clasp • Insert the tension strap into the tension strap clamp (2). • Place the tension strap around the pipe and insert it into the tension strap clamp on the opposite side and the tension strap, see Fig.
  • Page 38 6 Mounting 6.2 Transducers FLUXUS G722ST-HT Ratchet clasp • Insert the tension strap into the tension strap clamp (2) and the metal spring, see Fig. 6.23 and Fig. 6.26. • Place the tension strap around the pipe and insert it into the tension strap clamp on the opposite side and the slot of the clasp screw, see Fig.
  • Page 39 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Fig. 6.28: Pipe with 2 rails Step 5: Installation of the transducers in the mounting fixture Variofix L • Press the transducers firmly into their clamping fixtures in the covers until they are tightly fixed, see Fig. 6.29. Observe the correct orientation of the transducers.
  • Page 40 6 Mounting 6.2 Transducers FLUXUS G722ST-HT • Fix the transducer cables with the strain relief clamp to protect them from mechanical strain. Transducer cables with stainless steel conduit are fixed inside the housing, transducer cables with plastic conduit are fixed to the outside of the housing, see Fig.
  • Page 41 6 Mounting FLUXUS G722ST-HT 6.2 Transducers 6.2.2.5 Mounting with Variofix C Mounting For measurements in diagonal arrangement, the transducer mounting fixtures are mounted on opposite sides of the pipe. Depending on the transducer distance a and the measuring barrier distance Δx, 1 or 2 pairs of rails have to be mounted. The distances a and Δx are either summed up (1 pair of rails) or considered separately (2 pairs of rails) (see Tab.
  • Page 42 6 Mounting 6.2 Transducers FLUXUS G722ST-HT Step 1: Disassembly of the transducer mounting fixture Variofix C • Disassemble the transducer mounting fixture Variofix C. In order to remove the cover from the rail, bend the outer sides of the cover outwards. In order to remove the spring clip from the rail, slide it over the indentations on the rail and lift it off.
  • Page 43 6 Mounting FLUXUS G722ST-HT 6.2 Transducers • Insert approx. 100 mm of the tension strap into one of the slots of the tension strap clamp and bend it (see Fig. 6.36). • Insert the long end of the tension strap into the metal spring and the 2 tension strap clamps (see Fig. 6.37). Place the tension strap around the pipe (see Fig.
  • Page 44 6 Mounting 6.2 Transducers FLUXUS G722ST-HT Mounting the rail with ratchet clasp • Cut the tension strap to length (pipe circumference + at least 120 mm). Caution! The edge of the tension strap is very sharp. Risk of injury! → Debur sharp edges. →...
  • Page 45 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Fig. 6.42: Ratchet clasp with tension strap 1 – sense of rotation 2 – edge 3 – lever 4 – clasp screw with slot • Put the rails on the tension strap clamps (see Fig. 6.43). •...
  • Page 46 6 Mounting 6.2 Transducers FLUXUS G722ST-HT • Tighten the tensioning screws until the collar of the rivet nut is flush with the clip (see Fig. 6.45). • Fix the spacing elements on the rail to mark the transducer position. • Check whether the transducer distance of transducer pair B corresponds to the value displayed by the transmitter. •...
  • Page 47 6 Mounting FLUXUS G722ST-HT 6.2 Transducers Fig. 6.47: Removal of the cover 1 – lever tool 2 – fixture 6.2.2.6 Installation with WaveInjector The WaveInjector is a transducer mounting fixture for the ultrasonic flow measurement at extreme temperatures. Fig. 6.48: Transducer mounting fixture for WaveInjector For the installation of the WaveInjector, see document SU_WI-xxx.
  • Page 48: Temperature Probe

    6 Mounting 6.3 Temperature probe FLUXUS G722ST-HT Temperature probe Mounting 6.3.1 Pipe preparation Caution! Contact with grinding dust This may result in injuries (e.g., breathing difficulties, skin reactions, eye irritations). → Wear the required personal protective equipment. → Observe the applicable rules. Important! The pipe has to be sufficiently stable to withstand the pressure exerted by the temperature probe.
  • Page 49 6 Mounting FLUXUS G722ST-HT 6.3 Temperature probe Fig. 6.49: Clasp Fig. 6.50: Temperature probe on the pipe 6.3.2.2 Installation with FLEXIM clasp Caution! The edge of the tension strap is very sharp. Risk of injury! → Debur sharp edges. → Wear the required personal protective equipment. →...
  • Page 50 6 Mounting 6.3 Temperature probe FLUXUS G722ST-HT 6.3.2.3 Installation with quick release clasp Caution! The edge of the tension strap is very sharp. Risk of injury! → Debur sharp edges. → Wear the required personal protective equipment. → Observe the applicable rules. •...
  • Page 51: Connection

    7 Connection FLUXUS G722ST-HT Connection Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Installation, connection and start-up by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 52: Transducers

    7 Connection 7.1 Transducers FLUXUS G722ST-HT Transducers It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to avoid load on the connectors. Notice! If transducers are replaced, the SENSPROM has to be replaced as well. The transducers for the measuring barriers A and B have to be of the same type.
  • Page 53 7 Connection FLUXUS G722ST-HT 7.1 Transducers Fig. 7.2: Connection of the transducer cable with SMB connectors to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – SMB connector 2 – cable gland 7.1.1.2 Transducer cable with stainless steel conduit and stripped ends •...
  • Page 54 7 Connection 7.1 Transducers FLUXUS G722ST-HT Fig. 7.3: Connection of the transducer cable with stainless steel conduit and stripped ends to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – cable gland 7.1.2 Connection of the extension cable to the transmitter The extension cable is connected to the transmitter via the transducer connection.
  • Page 55 7 Connection FLUXUS G722ST-HT 7.1 Transducers Fig. 7.4: Connection of the extension cable to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE cable gland X_AV X_AR X_BV X_BR 1 – cap nut 2 – compression part 3 – basic part 4 –...
  • Page 56 7 Connection 7.1 Transducers FLUXUS G722ST-HT Fig. 7.5: Connection of the transducer cable with SMB connectors 1 – SMB connector 2 – cable gland 7.1.3.2 Transducer cable with stainless steel conduit and stripped ends • Remove the blind plug for the connection of the transducer cable. •...
  • Page 57 7 Connection FLUXUS G722ST-HT 7.1 Transducers Fig. 7.6: Connection of the transducer cable with stainless steel conduit and stripped ends 1 – cable gland 7.1.4 Connection of the extension cable to the junction box 7.1.4.1 Connection without potential separation (standard) The connection of the extension cable to the junction box without potential separation ensures that the transducer, junction box and transmitter are on the same potential.
  • Page 58 7 Connection 7.1 Transducers FLUXUS G722ST-HT Tab. 7.6: Terminal assignment terminal connection (extension cable) white or marked cable (core) white or marked cable (internal shield) brown cable (internal shield) brown cable (core) cable gland external shield Fig. 7.7: Connection of the extension cable to the junction box (without potential separation) cable gland 1 –...
  • Page 59 7 Connection FLUXUS G722ST-HT 7.1 Transducers 7.1.4.2 Connection with potential separation If earthing on the same potential cannot be ensured, e.g., in measurement arrangements with long extension cables, the extension cable and the junction box have to be electrically insulated from each other. The junction box and the transducers have to be on the same potential.
  • Page 60 7 Connection 7.1 Transducers FLUXUS G722ST-HT Fig. 7.8: Connection of the extension cable to the junction box (with potential separation) cable gland 1 – extension cable 2 – external shield 3 – shield terminal 4 – cap nut 5 – compression part 6 –...
  • Page 61 7 Connection FLUXUS G722ST-HT 7.1 Transducers 7.1.5 SENSPROM The SENSPROM contains important transducer data for the operation of the transmitter with the transducers. If transducers are replaced, the SENSPROM has to be replaced as well. Notice! The serial numbers of the SENSPROM and the transducer have to be identical. A wrong or wrongly connected SENSPROM will lead to incorrect measured values or to a measurement failure.
  • Page 62: Power Supply

    7 Connection 7.2 Power supply FLUXUS G722ST-HT Power supply Connection The installation of the power supply is carried out by the operator. The operator has to provide an overcurrent protector of max. 16 A (fuse or similar device) disconnecting all energizing wires in case of an inadmissible high current consumption. The impedance of the protective ground has to be low ohmic in order not to allow touch voltage pass the permissible limit.
  • Page 63: Cable Connection

    7 Connection FLUXUS G722ST-HT 7.2 Power supply Tab. 7.8: Terminal assignment terminal connection AC connection DC L(+) outer conductor N(-) neutral conductor protective conductor protective conductor fuse 1 A, time-lag 1.6 A, time-lag 7.2.1 Cable connection Transmitter with stainless steel housing •...
  • Page 64: Outputs

    7 Connection 7.3 Outputs FLUXUS G722ST-HT Transmitter with aluminum housing • Remove the blind plug to connect the cable to the transmitter. • Prepare the cable with a cable gland. The used cable has to have a wire cross-section of 0.25…2.5 mm •...
  • Page 65 7 Connection FLUXUS G722ST-HT 7.3 Outputs Tab. 7.9: Output circuits output transmitter external circuit remark internal circuit connection active current output/HART current output = 24 V < 500 Ω HART = 24 V < 500 Ω passive current output/HART current output = 8…30 V >...
  • Page 66 7 Connection 7.3 Outputs FLUXUS G722ST-HT Tab. 7.9: Output circuits output transmitter external circuit remark internal circuit connection digital output circuit 1 = 5…30 V < 100 mA [kΩ] = U [mA] circuit 2 RS485 120 Ω termination resistor shield The number, type and connections of the outputs depend on the order.
  • Page 67: Inputs

    7 Connection FLUXUS G722ST-HT 7.4 Inputs Inputs Important! The max. permissible voltage between the inputs and against PE is 60 V DC (permanent). Fig. 7.15: Connection of the inputs to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV...
  • Page 68 7 Connection 7.4 Inputs FLUXUS G722ST-HT Connection of a passive current sink Important! The terminals Txa and TxB must not be connected. Important! Observe the correct polarity in order to avoid damaging the current sink. A permanent short circuit can destroy the current input.
  • Page 69: Temperature Probe

    7 Connection FLUXUS G722ST-HT 7.5 Temperature probe Temperature probe Connection It is possible to connect the temperature probes Pt100/Pt1000 (4-wire) to the inputs of the transmitter (optional). Fig. 7.16: Connection of the temperature probes to the transmitter NEXT DISP SNAP DISP 3x Q FAST...
  • Page 70 7 Connection 7.5 Temperature probe FLUXUS G722ST-HT 7.5.1 Direct connection of the temperature probe • Remove the blind plug for the connection of the temperature probe. • Open the cable gland of the temperature probe. The compression part remains in the cap nut. •...
  • Page 71 7 Connection FLUXUS G722ST-HT 7.5 Temperature probe 7.5.2 Connection with extension cable • Remove the blind plug for the connection of the temperature probe. • Open the cable gland of the extension cable. The compression part remains in the cap nut. •...
  • Page 72 7 Connection 7.5 Temperature probe FLUXUS G722ST-HT Fig. 7.19: Preparation of the extension cable 110 mm 7 mm 1 – cap nut 2 – compression part 3 – basic part 4 – reducer 5 – external shield, brushed back 6 – sealing ring side Fig.
  • Page 73: Service Interfaces

    7 Connection FLUXUS G722ST-HT 7.6 Service interfaces Service interfaces Connection 7.6.1 USB interface The transmitter can be connected directly to the PC via the USB interface. • Connect the USB cable to the USB interface of the transmitter and to the PC. Fig.
  • Page 74 7 Connection 7.6 Service interfaces FLUXUS G722ST-HT Fig. 7.23: Cable gland 1 – cap nut 2 – compression part 3 – basic part 4 – sealing ring (only for cable gland M20, not for cable gland 1/2 NPS) 5 – housing wall 6 –...
  • Page 75 7 Connection FLUXUS G722ST-HT 7.6 Service interfaces Transmitter with aluminum housing Fig. 7.24: Connection of the LAN cable 1 – LAN interface 2 – foldable ferrite core 3 – LAN connector • Remove the blind plug to connect the cable to the transmitter. •...
  • Page 76: Start-Up

    8 Start-up FLUXUS G722ST-HT Start-up Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Installation, connection and start-up by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 77: Start-Up Settings

    8 Start-up FLUXUS G722ST-HT 8.1 Start-up settings Start-up settings When starting up the transmitter for the first time, the following settings are required: • language • time/date • system of units These displays will only be indicated when the transmitter is switched on for the first time or after an initialization. Language The available transmitter languages are displayed.
  • Page 78: Status Indications

    8 Start-up 8.3 Status indications FLUXUS G722ST-HT Status indications The operation state is indicated by LEDs. Fig. 8.1: Command panel of the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – LED (channel A) 2 –...
  • Page 79: Initialization

    8 Start-up FLUXUS G722ST-HT 8.5 Initialization Initialization During an initialization (INIT) of the transmitter, all settings are reset to the factory settings. Proceed as follows to execute an initialization: • When switching on a transmitter: keep and C pressed. • During the operation of the transmitter: press , C and ENTER simultaneously.
  • Page 80: Information Regarding The Transmitter

    8 Start-up 8.7 Information regarding the transmitter FLUXUS G722ST-HT Information regarding the transmitter Miscellaneous\System settings\Transmitter info • Select the menu item Transmitter info. • Press ENTER. • Press to scroll through the list. • Press to return to the menu item System settings. The following information relating to the transmitter is displayed: display description...
  • Page 81: Measurement

    9 Measurement FLUXUS G722ST-HT 9.1 Parameter input Measurement Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Caution! Warning of severe injuries from hot or cold components Touching hot or cold components can lead to severe injuries (e.g., thermal damage).
  • Page 82 9 Measurement 9.1 Parameter input FLUXUS G722ST-HT 9.1.1 Input of the measuring point number Parameters\Measuring point no. • Enter the number of the measuring point. • Press ENTER. For the activation of text input see Miscellaneous\Dialogs/Menus\Measuring point no. 9.1.2 Transducer selection Notice! The transducers have to be selected depending on the application parameters (see technical specification).
  • Page 83 9 Measurement FLUXUS G722ST-HT 9.1 Parameter input Pipe material Parameters\Pipe material The pipe material has to be selected to be able to determine the corresponding sound speed. The sound speeds for the materials in the scroll list are stored in the transmitter. •...
  • Page 84 9 Measurement 9.1 Parameter input FLUXUS G722ST-HT Lining material Parameters\Lining material • Select the lining material. • Press ENTER. • If the lining material is not included in the scroll list, select the list item Other material. • Press ENTER. This display will only be indicated if Yes is selected in the menu item Lining.
  • Page 85 9 Measurement FLUXUS G722ST-HT 9.1 Parameter input 9.1.4 Measurement settings Selection of the physical quantity Parameters\Physical quantity The available physical quantities are displayed in a list. • Select the physical quantity. • Press ENTER. Selection of the unit of measurement Parameters\Physical quantity\Flow velocity For the selected physical quantity, a scroll list with the available units of measurement is displayed.
  • Page 86 9 Measurement 9.1 Parameter input FLUXUS G722ST-HT Extension cable Parameters\Extension cable In case the transducer cable is extended (e.g., between junction box and transmitter), enter the length of the extension cable. • Enter the length of the extension cable. • Press ENTER. Input of the damping factor Parameters\Damping Each displayed measured value is a floating average of the last x seconds, with x being the damping factor.
  • Page 87: Configuration Of An Output

    9 Measurement FLUXUS G722ST-HT 9.2 Configuration of an output Input of the error delay Parameters\Error delay The error delay is the time interval after which the error value is transmitted to the output. This display will only be indicated if the list item Edit is selected in the menu item Miscellaneous\Dialogs/Menus\ Error delay.
  • Page 88: Fluid Properties

    9 Measurement 9.2 Configuration of an output FLUXUS G722ST-HT Tab. 9.1: Configuration of the outputs source item list item output Totalizers Volume (+) totalizer for the volumetric flow rate in positive flow direction Volume (-) totalizer for the volumetric flow rate in negative flow direction difference of the totalizers for the positive and negative flow direction Volume (Δ) totalizer for the mass flow rate in positive flow direction...
  • Page 89: Error Output

    9 Measurement FLUXUS G722ST-HT 9.2 Configuration of an output 9.2.1 Output of a measured value • Select the list item Outputs\...\Values. • Press ENTER. Source item Output range Values Outputs\...\Output range Output range • Select a list item. Error value –...
  • Page 90 9 Measurement 9.2 Configuration of an output FLUXUS G722ST-HT Tab. 9.3: Examples for the error output (output range: 4...20 mA) list item output signal Last value I [mA] 20.0 mA I [mA] Other value I [mA] error value = 3.5 mA Measuring range The sign of the measured value and the measuring range are determined.
  • Page 91: Terminal Assignment

    9 Measurement FLUXUS G722ST-HT 9.2 Configuration of an output Terminal assignment Outputs\...\Output info The terminals for the connection of the output are displayed. By pressing further information is displayed. • Press ENTER. Output function test The function of the output can now be tested. •...
  • Page 92 9 Measurement 9.2 Configuration of an output FLUXUS G722ST-HT 9.2.2 Output of a status/event value • Select the list item Outputs\...\Status. • Press ENTER. Source item Output range Status Outputs\...\Output range Output range • Select a list item. – 4...20 mA Status OK Idle state –...
  • Page 93 9 Measurement FLUXUS G722ST-HT 9.2 Configuration of an output Outputs\...\Test measuring range • Select Yes to test the status of the output signal. Select No to display the next menu item. • Press ENTER. Outputs\...\Enter test value • Select a list item as test value. •...
  • Page 94: Start Of Measurement

    9 Measurement 9.3 Start of measurement FLUXUS G722ST-HT Start of measurement Before starting the measurement, the measuring point has to be parameterized. If another measurement has already Installation Start measurement been performed using the same measuring point parameters, the measurement can be started immediately. Expected max.
  • Page 95: Display During The Measurement

    9 Measurement FLUXUS G722ST-HT 9.4 Display during the measurement Check of the transducer distance for transducer pair B Installation\...\Transducer distance The recommended transducer distance for transducer pair B will be displayed. • Check whether the transducer distance of transducer pair B corresponds to the value displayed by the transmitter. The transducer distances of transducer pair A and B have to be equal.
  • Page 96: Diagnostic Values

    9 Measurement 9.4 Display during the measurement FLUXUS G722ST-HT 9.4.2 Diagnostic values By pressing ENTER, the following diagnostic values can be displayed during the measurement: Tab. 9.4: Diagnostic values display description CrestFactor signal-to-noise ratio for the cross correlation function, reliability measure of the signal evaluation The measured value is valid if the crest factor is higher than the defined threshold value (default: crest factor >...
  • Page 97: Execution Of Special Functions

    9 Measurement FLUXUS G722ST-HT 9.5 Execution of special functions Notice! The parameters cannot be changed during the measurement. When attempting to change the parameters, the message Read-only mode will be displayed. The measurement has to be stopped in order to change the parameters. Information regarding the data logger Information regarding the data logger can be displayed during the measurement.
  • Page 98: Stop Of Measurement

    9 Measurement 9.6 Stop of measurement FLUXUS G722ST-HT Totalizers Measurement\Execute command\Totalizers • Select the list item Totalizers. • Press ENTER. The following scroll list appears: display description Reset totalizers reset the totalizer to zero Freeze display display the measured value of the totalizer for several seconds Reset error reset the totalizer error stop totalizer and reset it to zero...
  • Page 99: Troubleshooting

    10 Troubleshooting FLUXUS G722ST-HT Troubleshooting Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Service works by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 100: Problems With The Measurement

    10 Troubleshooting 10.1 Problems with the measurement FLUXUS G722ST-HT The display does not work at all or fails regularly. • Check the contrast setting of the transmitter or enter the HotCode 555000 to set the display to medium contrast. • Make sure that the correct voltage is available at the terminals. The destined transmitter voltage is indicated on the nameplate below the outer right terminal strip.
  • Page 101: Measuring Point Selection

    10 Troubleshooting FLUXUS G722ST-HT 10.2 Measuring point selection 10.2 Measuring point selection • Make sure that the recommended min. distance to any disturbance is observed. • Avoid measuring points with deposit formation in the pipe. • Avoid measuring points in the vicinity of deformations and defects on the pipe as well as welds. •...
  • Page 102: Maintenance And Cleaning

    11 Maintenance and cleaning FLUXUS G722ST-HT Maintenance and cleaning Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Service works by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 103: Maintenance

    11 Maintenance and cleaning FLUXUS G722ST-HT 11.1 Maintenance 11.1 Maintenance The transmitter and the transducers are practically maintenance-free. In order to ensure security, the following maintenance intervals are recommended: Tab. 11.1: Recommended maintenance intervals item maintenance step interval measure stainless steel housing visual inspection for corrosion annually or more cleaning...
  • Page 104: Dismounting And Disposal

    12 Dismounting and disposal 12.1 Dismounting FLUXUS G722ST-HT Dismounting and disposal Danger! Risk of explosion when using the measuring equipment in explosive atmospheres This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres". Warning! Installation, connection and start-up by unauthorized and unqualified personnel This may result in personal or material damage or dangerous situations.
  • Page 105: User Modes

    13 User modes FLUXUS G722ST-HT User modes The user modes allow extended diagnostics of signals and measured values as well as the definition of additional Miscellaneous parameters adapted to the application. The following user modes can be selected: System settings •StandardUser •ExpertUser User mode...
  • Page 106: Standarduser Mode

    13 User modes 13.1 StandardUser mode FLUXUS G722ST-HT 13.1 StandardUser mode In the StandardUser mode, all measurements can be carried out for the corresponding application. At the first start-up the transmitter operates in the StandardUser mode. 13.1.1 Parameter search range The transmitter calculates the expected transit time of the ultrasonic signal from the defined fluid sound speed.
  • Page 107 13 User modes FLUXUS G722ST-HT 13.1 StandardUser mode 13.1.4 Averaging factor of the measurement The averaging factor of the measurement defines a time range in which the values are averaged for calculation. A high factor increases the measurement stability, but also the response time to changed measured values. Miscellaneous\Special settings\Averaging factor •...
  • Page 108: Expertuser Mode

    13 User modes 13.2 ExpertUser mode FLUXUS G722ST-HT 13.1.7 Multi-point calibration It is possible to enter a series of measured values in order to define a calibration curve for the flow velocity. Record of a series of measured values: • Start a measurement with the transmitter and a reference flowmeter. •...
  • Page 109 13 User modes FLUXUS G722ST-HT 13.2 ExpertUser mode 13.2.1 Threshold value for valid measurement Miscellaneous\Special settings\Crest factor threshold • Select the menu item Special settings in the menu Miscellaneous. • Press ENTER until the menu item Crest factor threshold is displayed. •...
  • Page 110: Superuser Mode And Superuser Ext. Mode

    13 User modes 13.3 SuperUser mode and SuperUser ext. mode FLUXUS G722ST-HT Calibration\-Flow velocity limit • Enter a limit of the flow velocity for the measurement against the flow direction. • Press ENTER. If the flow velocity is lower than this limit, it will be marked as invalid. The physical quantity cannot be determined. UNDEF will be displayed.
  • Page 111: Calibration\Factor

    13 User modes FLUXUS G722ST-HT 13.3 SuperUser mode and SuperUser ext. mode 13.3.1 Linear calibration It is possible to define a correction of the flow velocity: = m · v + n where – measured flow velocity – factor, range: -2...+2 –...
  • Page 112: Outputs

    14 Outputs 14.1 Configuration of a digital output as binary output FLUXUS G722ST-HT Outputs If the transmitter is equipped with outputs, they have to be configured. For the configuration of the analog output, see Outputs section 9.2. The transmitter can also be equipped with digital outputs. Select output A digital output combines the functions of the following Enable output...
  • Page 113: Idle State

    14 Outputs FLUXUS G722ST-HT 14.1 Configuration of a digital output as binary output Assignment of a source item A source item has to be assigned to each selected output. Outputs\Source item Tab. 14.2: Output of status values or event values source item status value event value...
  • Page 114 14 Outputs 14.1 Configuration of a digital output as binary output FLUXUS G722ST-HT Output function test The function of the output can now be tested. • Connect a multimeter to the output. Outputs\...\B1 Test signal • Select Yes to test the output. Select No to display the next menu item. •...
  • Page 115: Configuration Of A Digital Output As Pulse Output

    14 Outputs FLUXUS G722ST-HT 14.2 Configuration of a digital output as pulse output Tab. 14.5: Output function test – measuring range list item switching function test value description Status OK Status OK • binary output is energized (status value) • measured value has to be low ohmic Status error •...
  • Page 116 14 Outputs 14.2 Configuration of a digital output as pulse output FLUXUS G722ST-HT Tab. 14.6: Selection of the physical quantity source item list item output Pulse Pulse |V| pulse without considering the sign of the volumetric flow rate Pulse +V pulse for positive measured values of the volumetric flow rate pulse for negative measured values of the volumetric flow rate Pulse -V...
  • Page 117 14 Outputs FLUXUS G722ST-HT 14.2 Configuration of a digital output as pulse output 14.2.2 Pulse output by defining pulses per unit Outputs\Pulse output • Select the list item Pulses per unit. • Press ENTER. Outputs\Output range • Select a list item: –...
  • Page 118: Configuration Of A Digital Output As Frequency Output

    14 Outputs 14.3 Configuration of a digital output as frequency output FLUXUS G722ST-HT Output function test Outputs\...\B1 Test signal • Select Yes to test the status of the output signal. Select No to display the next menu item. • Press ENTER. Outputs\...\B1 Enter test value •...
  • Page 119 14 Outputs FLUXUS G722ST-HT 14.3 Configuration of a digital output as frequency output Output range Outputs\...\Output range • Select a list item. – 0...1 kHz – 0...10 kHz – Other range • Press ENTER. If Other range is selected, enter the values Output MIN and Output MAX. Error output Outputs\...\Error value An error value can be defined which is output if the source item cannot be measured.
  • Page 120 14 Outputs 14.3 Configuration of a digital output as frequency output FLUXUS G722ST-HT Output function test The function of the output can now be tested. • Connect an external measuring instrument to the terminals of the installed output. Outputs\...\Test signal •...
  • Page 121: Inputs

    15 Inputs FLUXUS G722ST-HT 15.1 Configuration of an input Inputs Inputs Configure inputs Assign inputs → Current Ix Assignment of configured → Voltage Ux inputs → Temperature Tx see annex A 15.1 Configuration of an input If the transmitter is equipped with inputs, they have to be configured. •...
  • Page 122: Current And Voltage Inputs

    15 Inputs 15.1 Configuration of an input FLUXUS G722ST-HT 15.1.1 Current and voltage inputs When configuring the current or voltage inputs, the source item can now be selected and the input and measuring range is defined. Selection of the source item Inputs\...\Source item •...
  • Page 123: Temperature Inputs

    15 Inputs FLUXUS G722ST-HT 15.1 Configuration of an input 15.1.2 Temperature inputs When configuring a temperature input, the temperature probe can now be selected. Selection of the temperature probe Inputs\Configure inputs\Temperature Tx\Pt100/Pt1000 • Select the temperature probe: – Pt100 – Pt1000 Activation of the temperature correction A temperature correction (offset) can be set for each temperature input.
  • Page 124 15 Inputs 15.1 Configuration of an input FLUXUS G722ST-HT 15.1.3 Definition of a switching condition If a transmitter function is to be performed by remote control, a switching condition has to be defined. Inputs\...\Trigger value • Select Yes if a switching condition is to be defined. Select No to display the next menu item. •...
  • Page 125: Assignment Of An Input

    15 Inputs FLUXUS G722ST-HT 15.2 Assignment of an input 15.1.4 Terminal assignment Inputs\...\Input info The terminals for the connection of the input are displayed. By pressing further information is displayed. • Press ENTER. 15.1.5 Function test of the input The function of the installed input can now be tested. Analog input •...
  • Page 126: Data Logger

    16 Data logger 16.1 Configuration of the data logger FLUXUS G722ST-HT Data logger The transmitter has a data logger which stores the measured values during the measurement. Storage Data logger Notice! In order to store measured data, the data logger has Configuration to be configured.
  • Page 127 16 Data logger FLUXUS G722ST-HT 16.1 Configuration of the data logger display description Immediately The storing starts immediately. Full 5 minutes The storing starts in the next full 5 minutes. Full 10 minutes The storing starts in the next full 10 minutes. Full 15 minutes The storing starts in the next full 15 minutes.
  • Page 128 16 Data logger 16.1 Configuration of the data logger FLUXUS G722ST-HT Storage\Data logger\Configuration\Start storing\Event-based\Storage rate (trig.) • Select in the scroll list a storage rate for storing the measured values in case the event occurs. • Press ENTER. Storage\Data logger\Configuration\Start storing\Event-based\Buffer time ->П •...
  • Page 129: Deletion Of The Data Logger

    16 Data logger FLUXUS G722ST-HT 16.2 Deletion of the data logger Storage mode Storage\Data logger\Configuration\Storage mode • Select the menu item Storage\Data logger\Configuration. • Press ENTER until the menu item Storage mode is displayed. • Press ENTER. • Select Sample to store the current measured value. •...
  • Page 130: Information Regarding The Data Logger

    16 Data logger 16.3 Information regarding the data logger FLUXUS G722ST-HT 16.3 Information regarding the data logger Storage\Data logger\Data logger info • Select the menu item Storage\Data logger\Data logger info. • Press ENTER. The following information regarding the data logger is displayed: display description data logger is activated/deactivated...
  • Page 131: Data Transmission

    17 Data transmission FLUXUS G722ST-HT 17.1 Service interfaces Data transmission The data is transmit via the service interfaces or the process interface (optional) of the transmitter. 17.1 Service interfaces By means of the program FluxDiagReader it is possible to transmit via service interfaces (USB, LAN) data to the PC. The following tasks can be carried out: •...
  • Page 132: Process Interface

    17 Data transmission 17.2 Process interface FLUXUS G722ST-HT Automatic addressing with DHCP • Select Automatic to automatically identify the network parameters (IP address, subnet mask and standard gateway) via a DHCP server. Notice! The network parameters can only be automatically identified if the LAN supports DCHP. •...
  • Page 133: Advanced Functions

    18 Advanced functions FLUXUS G722ST-HT 18.1 Totalizers Advanced functions 18.1 Totalizers The total volume or total mass of the fluid at the measuring point can be determined. There are 2 totalizers, one for the positive and the other for the negative flow direction. The unit of measurement used for totalizing corresponds to the volume or mass unit selected for the physical quantity.
  • Page 134: Detection Of Long Measurement Failures

    18 Advanced functions 18.1 Totalizers FLUXUS G722ST-HT The number of decimal places is constant. The max. value of the totalizers decreases with an increasing number of decimal places. decimal places max. value max. display < 10 ±9999999999 < 10 ±99999999.9 <...
  • Page 135 18 Advanced functions FLUXUS G722ST-HT 18.1 Totalizers Notice! The overflow of a totalizer influences all output channels, e.g., the data logger and the online transmission of data. The sum of both totalizers (throughput Q) transmitted via an output is no longer valid after one of the totalizers has overflowed for the first time.
  • Page 136: Diagnosis With The Help Of The Snap Function

    18 Advanced functions 18.2 Diagnosis with the help of the snap function FLUXUS G722ST-HT 18.2 Diagnosis with the help of the snap function 18.2.1 Configuration By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of measuring results or for diagnostic purposes.
  • Page 137: Modification Of The Limit For The Inner Pipe Diameter

    18 Advanced functions FLUXUS G722ST-HT 18.3 Modification of the limit for the inner pipe diameter 18.2.4 Deletion of snaps Storage\Snap\Delete snaps • Select the menu item Snap\Delete snaps in the menu Storage. • Press ENTER. • Select Yes or No. •...
  • Page 138 18 Advanced functions 18.4 Remote functions FLUXUS G722ST-HT 18.4.1 Configuration of the remote function Functions\Remote functions • Select the menu item Remote functions in the menu Functions. • Press ENTER. In the scroll list of the functions it is displayed whether, and if so, which input or event trigger is assigned to a function. •...
  • Page 139: Event Triggers

    18 Advanced functions FLUXUS G722ST-HT 18.5 Event triggers 18.5 Event triggers It is possible to configure max. 4 independent event trigger R1, R2, R3, R4. Functions The event triggers can be used, e.g., to: Event trigger – output information about the running measurement –...
  • Page 140 18 Advanced functions 18.5 Event triggers FLUXUS G722ST-HT Tab. 18.2: Source items source item list item output Fluid properties Fluid temp. fluid temperature Fluid pressure fluid pressure fluid density Fluid density kinematic viscosity Kin. viscosity Dyn. viscosity dynamic viscosity Compress. coeff. gas compressibility coefficient Diagnostic values Crest factor...
  • Page 141 18 Advanced functions FLUXUS G722ST-HT 18.5 Event triggers Definition of trigger limits Functions\Event trigger\Rx Enable\Source item\...\Trigger value The limits are to be entered at which the event trigger has to switch. • Enter the upper limit MAX (x>limit). • Press ENTER. •...
  • Page 142 18 Advanced functions 18.5 Event triggers FLUXUS G722ST-HT 18.5.1 Apparent switching delay The measured values and totalizer values will be displayed rounded according to the set number of decimal places. The limits, however, will be compared to the non-rounded measured values. This might cause an apparent switching delay when the measured value changes marginally (less than the visualized decimal places).
  • Page 143: Event Log

    18 Advanced functions FLUXUS G722ST-HT 18.6 Event log Example R1 = 18.6 Event log If an error occurs, an error message will be displayed in the first line indicating the symbol . The error message can be displayed. Storage\Event log •...
  • Page 144: Settings

    19 Settings 19.1 Dialogs and menus FLUXUS G722ST-HT Settings 19.1 Dialogs and menus Miscellaneous\Dialogs/Menus • Select the menu item Dialogs/Menus in the menu Miscellaneous. • Press ENTER. Pipe circumference Miscellaneous\Dialogs/Menus\Pipe circumference • Press ENTER until the menu item Pipe circumference is displayed. •...
  • Page 145 19 Settings FLUXUS G722ST-HT 19.1 Dialogs and menus Error delay The error delay is the time after which an error value will be sent to an output if no valid measured values are available. Miscellaneous\Dialogs/Menus\Error delay • Press ENTER until the menu item Error delay is displayed. •...
  • Page 146: Measurement Settings

    19 Settings 19.2 Measurement settings FLUXUS G722ST-HT 19.2 Measurement settings Miscellaneous\Measurement\Measurement settings • Select the menu item Measurement in the menu Miscellaneous. • Press ENTER. • Select the menu item Measurement settings. • Press ENTER. Multi-point calibration A multi-point calibration allows a very precise output of measuring results. It is based on calibration curves of series of measured values Miscellaneous\Measurement\Measurement settings\Multi-point calibration •...
  • Page 147: Units Of Measurement

    19 Settings FLUXUS G722ST-HT 19.3 Units of measurement 19.3 Units of measurement It is possible to set the global units of measurement for length, temperature, pressure, sound speed, density and kinematic viscosity. Miscellaneous\Units of measurement • Select the menu item Units of measurement in the menu Miscellaneous. •...
  • Page 148: Working With Parameter Records

    19 Settings 19.5 Working with parameter records FLUXUS G722ST-HT 19.5 Working with parameter records 19.5.1 Introduction Parameter records are data sets that contain all information necessary to perform a certain measurement task: • pipe parameters • transducer parameters • fluid parameters •...
  • Page 149: Contrast Settings

    19 Settings FLUXUS G722ST-HT 19.6 Contrast settings 19.6 Contrast settings Miscellaneous\System settings\Display contrast • Select the menu item System settings in the menu Miscellaneous. • Press ENTER. • Select the menu item Display contrast. • Press ENTER. The display contrast is adjusted with the following keys: increases the contrast reduces the contrast •...
  • Page 150: Language Selection

    19 Settings 19.8 Key lock FLUXUS G722ST-HT Language selection The language selection can either be carried out via the menu Miscellaneous\System settings\Language or a HotCode: language HotCode English 909044 German 909049 French 909033 Spanish 909034 Dutch 909031 Russian 909007 Polish 909048 Turkish 909090...
  • Page 151 19 Settings FLUXUS G722ST-HT 19.8 Key lock Intervention in the measurement If the key lock is activated, the message Key lock activated will be displayed for a few seconds when pressing a key. In order to interrupt a measurement, the key lock has to be deactivated. •...
  • Page 152 19 Settings 19.8 Key lock FLUXUS G722ST-HT 2024-03-01, UMFLUXUS_G722ST-HTV2-2EN...
  • Page 153: Annex

    Annex FLUXUS G722ST-HT A Menu structure Annex Menu structure Main menu Parameters Installation Start measurement Inputs Outputs Measuring point no. Expected max. flow Measurement Select outputs → Configure inputs → Assign inputs Transducer Transducer distance A Enable Pipe parameters Barrier distance Source item Physical quantity Transducer distance B...
  • Page 154: Parameter Input

    Annex A Menu structure FLUXUS G722ST-HT Parameter input Main menu Parameters Measuring point no. Transducer Connected transd. Transducer not found Select transducer Outer diameter Pipe material Pipe wall thickness Lining Lining 2 Coating Roughness Automatic Customized Physical quantity Unit of measurement Fluid temp.
  • Page 155 Annex FLUXUS G722ST-HT A Menu structure Data logger Storage Data logger Configuration Delete meas. values Data logger info Enable logger Start storing → Ringbuffer activated → Logger full → Meas. val. Series → Occup. Logger Event-based → Immediately → Full 5 minutes →...
  • Page 156: Measurement Start

    Annex A Menu structure FLUXUS G722ST-HT Measurement start Installation Start measurement Main menu Expected max. flow Transducer distance A Barrier distance Transducer distance B Barrier distance Measurement Show measurement Stop measurement Main menu Read-only mode 2024-03-01, UMFLUXUS_G722ST-HTV2-2EN...
  • Page 157: Output Configuration

    Annex FLUXUS G722ST-HT A Menu structure Output configuration Outputs → Current Current output modes → Voltage → Digital output Active Passive Enable ... Source item Sound speed Flow quantities Totalizers Fluid properties Diagnostic values Miscellaneous Event trigger Pulse Pulse mode Status Values →...
  • Page 158 Annex A Menu structure FLUXUS G722ST-HT Inputs Inputs Assign inputs Configure inputs Available physical → Temperature quantities → Current → Voltage Available inputs Temperature Enable Tx Pt100/Pt1000 Temperature offset → Temperature Trigger value Function → Within range → MAX (x>limit) →...
  • Page 159 Annex FLUXUS G722ST-HT A Menu structure Inputs Inputs Configure inputs Current Enable Ix → Temperature → Current → Voltage Source item Customized Current → Temperature → Pressure → Density Input name → Kin. viscosity → Dyn. viscosity Unit of measurement Decimal places Input range Other range...
  • Page 160 Annex A Menu structure FLUXUS G722ST-HT Inputs Inputs Configure inputs Voltage Enable Ux → Temperature → Current → Voltage Source item Customized Current → Temperature → Pressure → Density Input name → Kin. viscosity → Dyn. viscosity Unit of measurement Decimal places Input range Other range...
  • Page 161: Special Settings

    Annex FLUXUS G722ST-HT A Menu structure Special settings Miscellaneous Special settings Signal search range High voltage Automatic Advanced search run Averaging factor Crest factor threshold Use satur. steam curve Tf->Pf Pf->Tf Legend only in ExpertUser, SuperUser and SuperUser ext. mode only with steam measurement UMFLUXUS_G722ST-HTV2-2EN, 2024-03-01...
  • Page 162 Annex A Menu structure FLUXUS G722ST-HT Calibration If Channel-based param. is selected for the configuration of the measuring point, the settings can be made individually for each measuring channel. Calibration Cut-off flow Flow velocity limit Linear calibration Profile correction Default Default Customized Multi-point calibration...
  • Page 163 Annex FLUXUS G722ST-HT A Menu structure Results – overview Trigger Input quantities Physical quantities Inputs Parameters → Temperature → Flow quantities → Current → Totalizers → Fluid properties → Diagnostic values Condition Event trigger → Switching condition → Holding behavior →...
  • Page 164 Annex A Menu structure FLUXUS G722ST-HT Definition of event triggers Functions Event trigger Select Rx Enable Rx Source item Sound speed Flow quantities Totalizers Fluid properties Diagnostic values Miscellaneous Function → MAX (x>limit) → MIN (x<limit) → Within range → Out of range →...
  • Page 165: Remote Functions

    Annex FLUXUS G722ST-HT A Menu structure Remote functions Trigger Inputs Functions Condition Event trigger Temperature Current Enable Enable Select Rx Pt100/Pt1000 Source item Enable Rx Source item Condition Trigger value Function → MAX (x>limit) → MIN (x<limit) → Within range →...
  • Page 166 Annex A Menu structure FLUXUS G722ST-HT Event-based storing of measured values Trigger Inputs Functions Condition Event trigger Temperature Current Enable Enable Select Rx Pt100/Pt1000 Source item Enable Rx Source item Condition Trigger value Function → MAX (x>limit) → MIN (x<limit) →...
  • Page 167: Units Of Measurement

    Annex FLUXUS G722ST-HT B Units of measurement Units of measurement Length/roughness unit of measurement description millimeter inch Temperature unit of measurement description °C degree Celsius degree Fahrenheit °F Pressure unit of measurement description bar(a) bar (absolute) bar(g) bar (relative) psi(a) pound per square inch (absolute) pound per square inch (relative) psi(g)
  • Page 168 Annex B Units of measurement FLUXUS G722ST-HT Operating volumetric flow rate unit of measurement description operating volume (totalized) m3/d cubic meter per day m3/h cubic meter per hour cubic meter per minute m3/min m3/s cubic meter per second km3/h cubic kilometer per hour ml/min milliliter per minute liter per hour...
  • Page 169 Annex FLUXUS G722ST-HT B Units of measurement unit of measurement description operating volume (totalized) Igpd (Imp-gal/d) gallon per day Igal Igph (Imp-gal/h) gallon per hour Igal Igpm (Imp-gal/m) gallon per minute Igal Igps (Imp-gal/s) gallon per second Igal imperial kilogallon per minute IKGM (Imp-Kgal/m) IMGD (Imp-Mgal/d) million imperial gallons per day...
  • Page 170: C Reference

    Annex C Reference FLUXUS G722ST-HT Reference The following tables provide assistance for the user. The accuracy of the data depends on the composition, temperature and processing of the material. FLEXIM does not assume liability for any inaccuracies. Sound speed of selected pipe and lining materials at 20 °C The values of some of these materials are stored in the internal database of the transmitter.
  • Page 171: D Legal Information - Open Source Licenses

    Annex FLUXUS G722ST-HT D Legal information – open source licenses Legal information – open source licenses The software of this product contains the following open-source software which is subject to the Apache license version 2.0 dated January 2004: 1. uC-TCP-IP https://github.com/weston-embedded/uC-TCP-IP/tree/v3.06.01 https://github.com/weston-embedded/uC-TCP-IP/blob/v3.06.01/LICENSE https://github.com/weston-embedded/uC-TCP-IP/blob/v3.06.01/NOTICE...
  • Page 172 Annex D Legal information – open source licenses FLUXUS G722ST-HT 3. uC-DHCPc https://github.com/weston-embedded/uC-DHCPc/tree/v2.11.01 https://github.com/weston-embedded/uC-DHCPc/blob/v2.11.01/LICENSE https://github.com/weston-embedded/uC-DHCPc/blob/v2.11.01/NOTICE ATTENTION ALL USERS OF THIS REPOSITORY: The original work found in this repository is provided by Silicon Labs under the Apache License, Version 2.0. Any third party may contribute derivative works to the original work in which modifications are clearly identified as being licensed under: (1) the Apache License, Version 2.0 or a compatible open source license;...
  • Page 173 Annex FLUXUS G722ST-HT D Legal information – open source licenses TERMS AND CONDITIONS FOR USE, REPRODUCTION, AND DISTRIBUTION 1. Definitions. "License" shall mean the terms and conditions for use, reproduction, and distribution as defined by Sections 1 through 9 of this document. "Licensor"...
  • Page 174 Annex D Legal information – open source licenses FLUXUS G722ST-HT (d) If the Work includes a "NOTICE" text file as part of its distribution, then any Derivative Works that You distribute must include a readable copy of the attribution notices contained within such NOTICE file, excluding those notices that do not pertain to any part of the Derivative Works, in at least one of the following places: within a NOTICE text file distributed as part of the Derivative Works;...
  • Page 175: Conformity Declarations

    Annex FLUXUS G722ST-HT E Conformity declarations Conformity declarations For EU declaration of conformity according to ATEX directive, see document SIFLUXUS. For the UK declaration of conformity according to the legal UK regulations, see document SI_UKx78xx. UMFLUXUS_G722ST-HTV2-2EN, 2024-03-01...
  • Page 176 Annex E Conformity declarations FLUXUS G722ST-HT 2024-03-01, UMFLUXUS_G722ST-HTV2-2EN...
  • Page 180 Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Flexim is a mark of one of the Emerson family of companies. All other marks are the property of their respective...