Error text
Sensor current too high
No standard
Calibration value sensor1(2) too low
Calibration value sensor1(2) too high
Explanation and rectification
Device was removed from the mains for some hours
or days
→ Chip sensor stabilizes. Wait until device
automatically goes into readiness to measure.
Chip sensor has been changed
→ Chip sensor stabilizes. Wait until device
automatically goes into readiness to measure.
→ Check for small air bubbles in the flow system
→ Perform Sensor conditioning again and select
new to perform the full conditioning cycle.
See Section 6.3.
Chip sensor defect or too old
→ Change chip sensor
System solution empty or unsuitable
→ Refill fresh system solution
At measurement start or calibration the device
cannot find any multi standard micro test tube, or
the multi standard micro test tube inserted is empty.
→ Insert multi standard micro test tube
→ Replace empty multi standard micro test tube
with a full one
→ Remove multi standard micro test tube briefly
and insert it again
After each action start measurement again.
Multi standard micro test tube is empty
→ Exchange multi standard micro test tube
Incorrect or unsuitable multi standard
→ Use a new original multi standard micro test tube,
if possible from new packaging
Chip sensor defect or too old
→ Change chip sensor
Leaky flow system. Exchanger, sample cannula or
pump tube worn out or defect
→ Control flow system, change exchanger, sample
cannula or pump tube if required
→ Perform Sensor conditioning again and select
new to perform the full conditioning cycle. See
Section 6.3.
When the sensor has been changed
→ Perform Sensor conditioning again and select
new to perform the full conditioning cycle. See
Section 6.3.
Incorrect or unsuitable multi standard
→ Use a new original multi standard micro test tube,
if possible from new packaging
Chip sensor defect or too old
→ Change chip sensor