Page 1
HEAT INDEX WBGT METER Model : WBGT-2009 Your purchase of this HEAT INDEX WBGT METER marks a step forward for you into the field of precision measurement. Although this meter is a complex and delicate instrument, its durable structure will...
Page 2
TABLE OF CONTENTS 1. FEATURES............................. 2. SPECIFICATIONS.......................... 3. FRONT PANEL DESCRIPTION..................... 3-1 Display............................. 3-2 Power Button ( ESC, Backlight Button ).................. 3-3 Hold Button ( Function, Next Button )..................3-4 REC Button ( Enter Button, in/Outdoor Button )..............3-5 SET Button ( ▼ Button )................3-6 ▲...
Page 3
1. FEATURES * Introduction : WBGT ( wet-bulb globe temperature ) was developed as a monitoring basis at US military training camps and became widespread for the use in workplaces and sports situation It is suggested in the international standard ISO 7243, OSHA ( US Occupation Safety ), Japan Society for Occupational Health and the SMA ( Sports Medicine Australia ) in establishing the permissible heat exposure limits in health , sports or...
2. SPECIFICATIONS 2-1 General Specifications Circuit Custom one-chip of microprocessor LSI circuit. Display LCD size : 52 mm x 38 mm LCD with green backlight ( ON/OFF ). Measurement * Wet bulb globe temperature ( WBGT ) @ Indoor/Outdoor * Black globe temperature ( TG ) * Air temperature ( TA ) * Wet bulb temperature ( WB ) * Humidity...
Page 5
Data Output RS 232/USB PC computer interface. Connect the optional RS232 cable UPCB-02 will get the RS232 plug. Connect the optional USB cable USB-01 will get the USB plug. Operating 0 to 50 ℃. Temperature Operating Less than 85% R.H. Humidity Power Supply * AC/DC power adapter is optional.
Page 6
2-2 Electrical Specifications (23±5 ℃ Wet bulb globe temperature ( WBGT ) Range Indoor 0 ℃ to 59 ℃ ℃ Outdoor 0 ℃ to 56 ℃ Indoor 32 to 138 ℉ ℉ Outdoor 32 to 132 ℉ Accuracy ±1 ℃ ( 15 to 59 ℃ ) ℃...
Page 7
Black globe temperature ( TG ) Range 0 to 80 ℃ ℃ 32 to 176 ℉. ℉ Resolution 0.1 ℃ ℃ 0.1 ℉ ℉ Accuracy ±0.6 ℃ ℃ ±1.1 ℉ @ 15 to 40 ℃ ℉ Humidity Range 5 % to 95 % R.H. Resolution 0.1 % R.H.
4. MEASURING PROCEDURE 4-1 Function selection 1) Turn on the meter by pressing the " Power Button " ( 3-2, Fig. 1 ) momentarily. Pressing the " Power Button " ( 3-2, Fig. 1 ) continuously and > 2 seconds again will turn off the meter.
Page 10
The meter Temp. display unit is defaulted to " ℃ " If intend to let the meter's temperature unit default to " ℉ " , then please refer chapter 5-4, page 12 ). Wet bulb globe Temp. ( WBGT ) measurement 1) Power on the meter by pressing the "...
Page 11
Dew point Temp. measurement 1) Power on the meter by pressing the " Power Button " ( 3-2, Fig. 1 ). 2) Function select to " Dew point Temp. " measurement. * The upper Display will show the Dew point Temp. value. The lower Display will show the text "...
Page 12
4-3 Data Record ( Max., Min. reading ) 1) The data record function records the maximum and minimum readings. Press the " REC Button " ( 3-4, Fig. 1 ) once to start the Data Record function and there will be a " REC " symbol on the display. 2) With the "...
5. ADVANCED SETTING Under do not execute the Datalogger function, press the " SET Button " ( 3-5, Fig. 1 ) continuously at least two seconds will enter the " Advanced Setting " mode. then press the " Next Button " (3-3, Fig. 1 ) once a while in sequence to select the eight main function, the lower display will show : AL...
Page 14
5-2 Auto power OFF management When the lower display show " PoFF " 1) Use the " ▲ Button " ( 3-6, Fig. 1 ) or " ▼ Button " ( 3-5, Fig. 1 ) to select the upper value to " yES " or "...
C - Temperature unit is ℃ F - Temperature unit is ℉ 2) After Display unit is selected to " C " or " F ", press the " Enter Button " ( 3-4, Fig. 1 ) will save the setting function with default.
8. SYSTEM RESET If the meter happen the troubles such as : CPU system is hold ( for example, the key button can not be operated... ). Then make the system RESET will fix the problem. The system RESET procedures will be either following method : During the power on, use a pin to press the "...
Page 17
Meter (9W 'D" Connector) Center Pin..........Pin 4 (3.5 mm jack plug) Ground/shield........... Pin 2 2.2 K resistor Pin 5 The 16 digits data stream will be displayed in the following format : D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status : Start Word When send the WBGT display data = 1...
Page 18
RS232 FORMAT : 9600, N, 8, 1 Baud rate 9600 Parity No parity Data bit no. 8 Data bits Stop bit 1 Stop bit 10. APPENDIX Recommendations to heat exposure level for working and physical activity. ( Permissible heat exposure threshold limit ) Work/rest regimen Light Moderate...
Page 19
WBGT INDEX and PHYSICAL EXERCISE WBGT Flag Activity Intensity < 26.7 ℃ White Normal activity. ( < 80 ℉ ) Caution should be taken. 26.7 to 29.4 ℃ Green Descretion is required ( 80 to 84.9 ℉ ) in planning intense physical activity.