Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code...
Page 6
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 1 User instructions WARNING In cases which – if not prevented – can lead to death or severe injury. CAUTION In cases which –...
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 2 Safety | 2.4 Accessories and combination with other equipment ▪ Damage (e.g., from dropping) ▪ Irregular running noise ▪ Excessive vibration ▪...
Page 9
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 2 Safety | 2.6 Service and repair 4 Have the service and testing tasks carried out according to the Medical Device Operator Ordinance.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 3 Product description | 3.2 Technical specifications S31, S31 K (without light) ▪ the applicable accident prevention regulations ▪ these Instructions for use According to these regulations, the user is required: ▪...
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 4 Startup and shut down 4 Startup and shut down WARNING Hazard from contaminated products. Patients, users or third parties could be infected by contaminated medical devices.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING The patient swallows or aspirates loose parts or substances.
Page 15
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 5 Operation | 5.3 Inserting the dental bur NOTE Only use prophy cups and polishing brushes or supports with Snap-On func- tion in accordance with DIN EN ISO 13295, support type 5.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device.
Page 18
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 7.1 Preparations at the site of use...
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual reprocessing CAUTION Never reprocess this medical device in an ultrasonic cleaner.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated reprocessing 4 Allow the disinfectant to act for 5 minutes.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing NOTE Please comply with the instructions for use of the washer disinfector.
Instructions for use SMARTmatic PROPHY S31 – 1.011.6760, SMARTmatic PROPHY S31 K – 1.011.6761, SMART- matic PROPHY S33 – 1.011.6800 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.6 Sterilisation 7.6 Sterilisation Sterilisation in a steam steriliser (autoclave) in...