Denotes conformity to specific safety standards and regulations. FDA LISTED FDA Listed Denotes that the product has been properly listed with the FDA. Denotes conformity to quality standards and quality management ISO Certification systems. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 1...
Electrical Safety protection is provided by properly connecting the centrifuge to earth ground. Use only the manufacturer provided line cord and ensure that it is connected to a properly grounded power receptacle. Failure to do so will result in an electrical hazard. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 2...
Page 5
WARNING: Do not make modifications to or remove any hardware from rotor without prior authorization from Drucker Diagnostics. WARNING: Only use Drucker Diagnostics components in this centrifuge. Due to the lack of the possibility of human exposure, all Drucker centrifuges and accessories sold by Drucker Diagnostics, Inc.
General purpose laboratory centrifuge, intended for the density-based separation of fluids through centripetal acceleration. ARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
Remove and discard label that is covering the START button. BE SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY DISCONNECTION! Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 5...
To re-enable cycle selection, open lid, then press and hold the UNLOCK Unlock Cycle button for 5 seconds. Three beeps will confirm that cycle selection is Selection now unlocked. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 6...
When 2 min cycle LED and Error/Service Indicator flash, it indicates cover opened or cycle changed during operation. When 30s cycle LED and Error/Service Indicator flash, it indicates the centrifuge unable to reach target rotor speed. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 7...
4. It is recommended that the initial setup procedures be performed to ensure that the rotor has been installed correctly and that no damage has been done to the centrifuge during either the rotor installation or possible rotor chamber cleaning. See page 4 for this procedure. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 8...
If no LED light is on, the unit is not powered, and the lid will not unlock by conventional means. Insert the provided tool through Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 10...
Page 13
Contact Nuo Therapeutics Customer Service for assistance. Clicking noise during braking gets loud o Make sure that the screw in the center of the rotor is tight. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 11...
50% relative humidity at 40 °C Overvoltage Category Pollution Degree *The rotor and accessories are rated for the maximum rotation frequency shown. Maximum sample density is 1.15 grams/mL, (water density = 1.0 grams/mL) Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 12...
Pediatric Tube Adapter, Stat Rotor, Printed (Pack of 6) 30-000212 Hex Key Use only approved accessories from the original manufacturer. A complete list of accessories is available at www.druckerdiagnostics.com. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 13...
Page 16
This operator’s manual is part number 03-0-0002-0315 Rev. A Product Family: DASH Series (DASH Micro) Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020 Protected by U.S. Patents #6,811,531, # 7,422,554, #D718,463, & #D734,489. Other Patents Pending FDA LISTED INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste.