IMPORTANT SAFETY INFORMATION

This section contains important safety information
regarding the power cell. For more safety information,
refer to the safety manual and the user guide supplied
with the hearing instruments.

DEFECTIVE POWER CELLS

In case you think the power cell is defective, comply with
the following steps:
X
X
If any of the test steps failed, consider the power cell 
defective. Due to safety concerns the hearing instrument 
6
RX_BiCore_Custom_Li_ITC_Dispenser_EN.indd 6
RX_BiCore_Custom_Li_ITC_Dispenser_EN.indd 6
NOTICE
Do not exchange power cells. This may only be
X
performed by Service Center.
Place the hearing instrument into the charger. If you 
get a red LED, the power cell is defective.
Perform visual inspection of the hearing instrument. 
The test is considered failed if it has any of below 
criteria:
- There is physical sign of deformation caused by
bloated lithium power cell inside.
- There is visible leakage from the lithium power cell.
- Product is generating heat on its own or hot to the
touch. It is however normal that the hearing instrument 
heats up when being charged.